TINA-QUANT ALBUMIN GEN.2 - URINE APPLICATION
Received Jul 22, 2025 · Event occurred Jul 1, 2025
Report 1823260-2025-02231 · MDR key 22563902
Device
Generic name
Albumin Immunological Test System
Manufacturer
Roche DiagnosticsCatalog number
04469658190
Lot number
845956
Product problems
- Low Test Results
- Infusion or Flow Problem
- Low Test Results
- Infusion or Flow Problem
Patient
67 YR · Male
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THE FIELD SERVICE ENGINEER (FSE) CHECKED THE PUMP PRESSURE, WATER PUMP, GEAR PUMP, AND VALVES. ALL WERE OPERATING WITH NO ISSUES. THE FSE PERFORMED SUCCESSFUL CALIBRATION, HARDWARE CHECKS, AND PRECISION. THE INSTRUMENT PERFORMANCE WAS VERIFIED. NO FURTHER ISSUES WERE REPORTED SINCE THE SERVICE VISIT. THE INVESTIGATION DID NOT IDENTIFY A PRODUCT PROBLEM. THE EXACT ROOT CAUSE COULD NOT BE DETERMINED. MEDWATCH FIELDS D1, D2, D4, G1, AND G4 HAVE BEEN UPDATED.
Additional Manufacturer Narrative
THE REAGENT LOT NUMBER WAS 845956. THE EXPIRATION DATE WAS REQUESTED, BUT NOT PROVIDED. THE INVESTIGATION IS ONGOING.
Description of Event or Problem
THERE WAS AN ALLEGATION OF QUESTIONABLE ALBT2 TINA-QUANT ALBUMIN GEN.2 FOR URINE SAMPLES FROM ONE PATIENT USING A COBAS 8000 COBAS C 502 MODULE. THE INITIAL RESULT WAS <12 MG/L WITH A DATA FLAG. THE FIRST REPEAT RESULT WAS 143 MG/L. THE SECOND AND THIRD REPEAT RESULTS WERE 44 MG/L AND 83 MG/L. THE FIRST REPEAT RESULT WAS QUESTIONED AS IT WAS CONSIDERED TOO HIGH WHICH TRIGGERED THE FURTHER REPEAT TESTING OF THE SAMPLE. NONE OF THE RESULTS WERE REPORTED OUT. IT IS UNKNOWN WHICH RESULTS WERE DEEMED CORRECT.