ATTUNE NOTCH GUIDE SZ 5
Received Jul 17, 2025 · Event occurred Jul 1, 2025
Report 1818910-2025-12411 · MDR key 22534028
Device
Generic name
Resection Blocks/devices
Manufacturer
Depuy Ireland - 9616671Catalog number
254500025
Lot number
PG239001
Product problems
- Scratched Material
- Scratched Material
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
PRODUCT COMPLAINT # (B)(4). INVESTIGATION SUMMARY - ACCORDING TO THE INFORMATION RECEIVED, "IT WAS REPORTED THAT DID NOT HAPPEN IN SURGERY, INSTRUMENTS ARE SCRATCHED AND DAMAGED.". THE PRODUCT WAS RETURNED TO J&J MEDTECH ORTHOPAEDICS FOR EVALUATION. VISUAL INSPECTION OF THE RETURNED DEVICE SHOWS SIGNS OF WEAR AND SCRATCHES THAT ARE CONSISTENT WITH CONSTANT USE, SERVICING, AND REPETITIVE STERILIZATION CYCLES. ADDITIONALLY, THE WHITE INK BACKFILLED INTO THE MARKINGS ON THE DEVICE WERE FADED. THE OBSERVED CONDITION WAS IDENTIFIED AS AN END-OF-LIFE INDICATOR, DAMAGE CONSISTENT WITH REPEATED USE AND SERVICING. THE LIFECYCLE REQUIREMENTS OF THE DEVICE ARE EVENT RELATED AND DEPEND ON THE USE AND INSPECTION OF THE DEVICE IN CLINICAL PRACTICE. AS THE DEVICE CAN BE DAMAGED ON THE FIRST OR 100TH USE, THE DEVICE MUST BE PROPERLY INSPECTED PRIOR TO EACH SURGICAL USE. REFER TO THE DEVICE/COUNTRY SPECIFIC IFU FOR INFORMATION RELATED TO END OF LIFE, REPROCESSING INSTRUCTIONS, AND INSPECTION PROCEDURES. A FUNCTIONAL TEST WAS NOT PERFORMED SINCE IT WAS NOT APPLICABLE TO THE COMPLAINT CONDITION. A DIMENSIONAL INSPECTION WAS NOT PERFORMED SINCE IT WAS NOT APPLICABLE TO THE COMPLAINT CONDITION. THE OVERALL COMPLAINT WAS CONFIRMED AS THE OBSERVED CONDITION OF THE ATTUNE NOTCH GUIDE WOULD HAVE CONTRIBUTED TO A DEVICE ISSUE. BASED ON THE INVESTIGATION FINDINGS, THE POTENTIAL CAUSE IS TRACED TO END OF LIFE, AND IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS REQUIRED. THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE REPORTED COMPLAINT CONDITION. AS PART OF J&J MEDTECH ORTHOPAEDICS QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. DEVICE HISTORY LOT - THE PRODUCT INVESTIGATION FOUND NO EVIDENCE SUSPECTING AN ERROR IN THE MANUFACTURING OR MATERIAL THAT WOULD BE A CONTRIBUTING FACTOR IN THE REPORTED ALLEGATION(S). A MANUFACTURING RECORDS EVALUATION (MRE) WAS NOT PERFORMED. H11 ADDITIONAL NARRATIVE: CORRECTED: H3.
Description of Event or Problem
IT WAS REPORTED THE INSTRUMENTS ARE SCRATCHED AND DAMAGED. EVENT DID NOT HAPPEN IN SURGERY.
Additional Manufacturer Narrative
THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803 (AND/OR PART 4, AS APPLICABLE). THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. ADDITIONAL NARRATIVE: IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
Additional Manufacturer Narrative
PRODUCT COMPLAINT # (B)(4). THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION THAT HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES JOINT RECONSTRUCTION, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES JOINT RECONSTRUCTION, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. H11 ADDITIONAL NARRATIVE: ADDED: D9. IF INFORMATION THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT IS OBTAINED, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.