inforMED
MalfunctionJCF

BD VACUTAINER® CPT¿ NH: ~130 IU FICOLL¿: 2.0ML

Received Jul 16, 2025 · Event occurred Jun 18, 2025

Report 1917413-2025-00930 · MDR key 22526828

Device

Generic name

Lymphocyte Separation Medium

Catalog number

362780

Lot number

UNKNOWN

Product problems

  • Incorrect, Inadequate or Imprecise Result or Readings
  • Device Ingredient or Reagent Problem
  • Incorrect, Inadequate or Imprecise Result or Readings
  • Device Ingredient or Reagent Problem

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

IT WAS REPORTED WHILE USING BD VACUTAINER® CPT¿ NH: ~130 IU FICOLL¿: 2.0ML, AN UNSPECIFIED NUMBER OF SAMPLES EXHIBIT OIL GEL GLOBULES AND LOW YIELD. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

Additional Manufacturer Narrative

B3: DATE OF EVENT IS UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. D4. MEDICAL DEVICE EXPIRATION DATE: UNKNOWN. H4. DEVICE MANUFACTURE DATE: UNKNOWN. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative

INVESTIGATION SUMMARY: BD RECEIVED 1 PHOTO FOR INVESTIGATION. THE PHOTO WAS REVIEWED AND THE INDICATED FAILURE MODE FOR OIL GEL GLOBULES WAS OBSERVED. LOW YIELD CANNOT BE SEEN FROM PHOTOS. BD WAS UNABLE TO DETERMINE THE SPECIFIC LOT NUMBER ASSOCIATED WITH THIS COMPLAINT; THEREFORE, A REVIEW OF THE DEVICE HISTORY RECORD COULD NOT BE CONDUCTED. COMPLAINTS FOR SAMPLE QUALITY ARE UNDER STATISTICAL CONTROL FOR THE MONTH OF JUNE 2025. AT THIS TIME, FURTHER TESTING IS NOT INDICATED. THIS COMPLAINT HAS BEEN CONFIRMED FOR THE INDICATED FAILURE MODE [OIL GEL GLOBULES BASED ON THE PHOTOS PROVIDED. BD WAS UNABLE TO CONFIRM THIS COMPLAINT FOR LOW YIELD. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS. BD QUALITY WILL CONTINUE TO MONITOR SAMPLE QUALITY COMPLAINTS.

Description of Event or Problem

IT WAS REPORTED WHILE USING BD VACUTAINER® CPT¿ NH: ~130 IU FICOLL¿: 2.0ML, AN UNSPECIFIED NUMBER OF SAMPLES EXHIBIT OIL GEL GLOBULES AND LOW YIELD. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.