RIGIDFIX FEM2.7MM B/T/B XPIN
Received Jul 15, 2025 · Event occurred Jun 25, 2025
Report 1221934-2025-02447 · MDR key 22506423
Device
Product problems
- Break
- Break
Patient
NA · Unknown
- Foreign Body In Patient
- Foreign Body In Patient
Narrative
Additional Manufacturer Narrative
THIS REPORT IS BEING SUBMITTED IN PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY MITEK OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, MITEK, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. H11 ADDITIONAL NARRATIVE: AS OF THIS DATE, THE DEVICE HAS NOT BEEN RETURNED FOR EVALUATION; THEREFORE, THE REPORTED CONDITION CANNOT BE CONFIRMED AND/OR DUPLICATED.
Description of Event or Problem
IT WAS REPORTED THAT DURING AN UNKNOWN SURGERY, IT WAS OBSERVED THAT THE RIGIDFIX FEM2.7MM B/T/B XPIN DEVICE SLEEVE ASSEMBLY WAS BROKEN WHILE IN USE. ACCORDING TO THE REPORTER THE PIN WAS EMBEDDED, DIFFICULT TO REMOVE AND APPROXIMATELY 1.5 CM OF METAL WAS LEFT IN SITU. IT WAS REPORTED THAT THE PIN WAS EMBEDDED AND THERE SHOULD NOT BE ANY PROBLEMS WITH IT. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Additional Manufacturer Narrative
THIS REPORT IS BEING SUBMITTED IN PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY MITEK OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, MITEK, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. H11 ADDITIONAL NARRATIVE: INVESTIGATION SUMMARY: THE COMPLAINT DEVICE WAS NOT RETURNED TO J&J MEDTECH ORTHOPAEDICS, THEREFORE UNAVAILABLE FOR A PHYSICAL EVALUATION. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE AND NO NON-CONFORMANCE WAS IDENTIFIED. BASED ON THE CURRENT AVAILABLE INFORMATION, WE CANNOT DETERMINE A ROOT CAUSE FOR THE REPORTED FAILURE. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, WE WILL REOPEN THE COMPLAINT AND PERFORM THE INVESTIGATION AS APPROPRIATE. IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED AND THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE REPORTED COMPLAINT CONDITION. AS PART OF J&J MEDTECH ORTHOPAEDICS QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES.