ENDO DISSECT
Received Jul 15, 2025 · Event occurred Jun 20, 2025
Report 2647580-2025-02114 · MDR key 22505074
Device
Product problems
- Detachment of Device or Device Component
- Detachment of Device or Device Component
Patient
NA · Female
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
Description of Event or Problem
IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY, THE ELECTRODE SEPARATED FROM THE BACK OF THE INSTRUMENT AT THE MOMENT THE DOCTOR PASSED THE DEVICE TO THE INSTRUMENTALIST. THERE WAS NO ELECTRODE BREAKAGE AND NO DEVICE COMPONENT FELL INTO THE CAVITY OF THE PATIENT. THE DOCTOR REPOSITIONED THE POSTERIOR ELECTRODE IN THE CLAMP TO RESOLVE THE ISSUE AND COMPLETE THE CASE. THERE WAS NO PATIENT INJURY.
Additional Manufacturer Narrative
ADDITIONAL INFORMATION: G3, H3, H6 H3 EVALUATION SUMMARY: MEDTRONIC CONDUCTED AN INVESTIGATION BASED UPON ALL INFORMATION RECEIVED. THE DEVICE WAS NOT RETURNED, BUT A PHOTO WAS AVAILABLE FOR EVALUATION. VISUAL INSPECTION NOTED THE FIRST IMAGE WAS OF THE DEVICE WITH A GREEN CIRCLE DRAWN AROUND THE PROXIMAL END OF THE HANDLE. THE QUALITY OF THE IMAGE WAS BLURRY. THE SECOND IMAGE DEPICTED THE SURGEON HOLDING THE DISTAL END OF THE DEVICE; THE ELECTRODE AT THE PROXIMAL END OF THE HANDLE WAS MISSING FROM VIEW. IT WAS REPORTED THAT THE ELECTRODE SEPARATED FROM THE BACK OF THE INSTRUMENT. THE REPORTED ISSUE WAS CONFIRMED. THE MOST LIKELY CAUSE COULD NOT BE ESTABLISHED FROM THE INFORMATION AVAILABLE. THE MANUFACTURING RECORDS FOR EACH DEVICE WERE THOROUGHLY REVIEWED PRIOR TO RELEASE TO ENSURE THAT IT MET ALL MEDTRONIC QUALITY SPECIFICATIONS. A SECONDARY REVIEW OF THE DEVICE HISTORY RECORDS FOUND NO POTENTIALLY CONTRIBUTING FACTORS. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.