ROBOTIC/LAPARASCOPY KITTNER ROLLS MEDIUM 4" LONG 5PK 25PKS/CS
Received Jul 10, 2025 · Event occurred May 19, 2025
Report 2951379-2025-00002 · MDR key 22469677
Device
Generic name
Sterile Kittner Roll Gauze
Manufacturer
Carefree Surgical Specialties, IncModel number
15505/25Lot number
1650
Product problems
- Manufacturing, Packaging or Shipping Problem
- Manufacturing, Packaging or Shipping Problem
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
CUSTOMER REPORTED INCORRECT QUANTITY IN SKU 15505/25 LOT: 1650. PACKAGE LABELED AS 5 PACK CONTAINED 6 UNITS. THE PRODUCT WAS NOT USED, AND NO PATIENT CONTACT OCCURRED. ROOT CAUSE DETERMINED TO BE OPERATOR OVERSIGHT DURING FINAL PACKAGING. STAFF WAS RETRAINED AND ADDITIONAL PACKAGIN CHECKS IMPLEMENTED. A REPLACEMENT PRODUCT WAS SENT. DEVICE WAS NOT RETURNED FOR EVALUATION. A VOLUNTARY RECALL WILL BE INITIATED IN JULY 2025.
Description of Event or Problem
A RECENT PACK OF KITTNER ROLLS SKU 15505/25 OPENED IN SURGERY WAS FOUND TO HAVE 6 ROLLS INSTEAD OF 5. THE OR STAFF DECIDED NOT TO USE THAT ONE DURING SURGERY IN CASE THERE WAS A DEFECT. THE OUTSIDE PACKAGING WAS DISCARDED SO LOT NUMBER WAS NOT AVAILABLE.
Remedial action
- Replace