HEXALUX
Received Jul 9, 2025 · Event occurred Jun 9, 2025
Report 1043572-2025-00062 · MDR key 22461877
Device
Generic name
Hexalux Ceiling Mount
Manufacturer
Steris CorporationModel number
HEXALUX CEILING MOUNTCatalog number
EXLCEIL
Product problems
- Detachment of Device or Device Component
- Detachment of Device or Device Component
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
A STERIS SERVICE INSTALLATION SPECIALIST ARRIVED ONSITE TO INSPECT THE HEXALUX CEILING-MOUNTED SURGICAL LIGHT. THE TECHNICIAN IDENTIFIED THE CAUSE OF THE REPORTED EVENT TO BE THE SNAP RING THAT SECURES THE LIGHT'S SUSPENSION ARM TO THE CEILING MOUNT HAD BEEN DAMAGED. THE CAUSE OF THE DAMAGE COULD NOT BE DETERMINED. AS THE SNAP RING WAS DAMAGED, OVER TIME, THIS ALLOWED FOR THE LIGHT'S SUSPENSION ARM TO LOOSEN AND EVENTUALLY DETACH, RESULTING IN THE REPORTED EVENT. THE USER FACILITY ELECTED TO HAVE THEIR HEXALUX CEILING-MOUNTED SURGICAL LIGHT REMOVED FROM SERVICE AND NOT REPAIRED OR REPLACED. NO ADDITIONAL ISSUES HAVE BEEN REPORTED.
Description of Event or Problem
THE USER FACILITY REPORTED THAT THEIR HEXALUX EXAMINATION LIGHT DETACHED FROM THE CEILING. NO REPORT OF INJURY.
Additional Manufacturer Narrative
OUR INVESTIGATION INTO THE REPORTED EVENT IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE. UDI RELATED DATA CLARIFICATION - THE LOT/SERIAL NUMBER IS UNKNOWN; THEREFORE, THE APPLICABLE PRODUCTION IDENTIFIERS WERE NOT INCLUDED IN THE MDR.