inforMED
MalfunctionFBO

CYSTO-URETHRO-FIBERSCOPE

Received Jul 8, 2025

Report 9610617-2025-01140 · MDR key 22439546

Device

Generic name

Cysto-urethro-fiberscope

Model number

8404ZXK

Catalog number

8404ZXK

Product problems

  • No Display/Image
  • Insufficient Device Problem Information
  • No Display/Image
  • Insufficient Device Problem Information

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE AFFECTED DEVICE HAS BEEN REQUESTED FOR INVESTIGATION BY THE MANUFACTURER. DEVICE WAS NOT RETURNED AT THE TIME OF MDR SUBMISSION. THE EVENT IS FILED UNDER INTERNAL KARL STORZ COMPLAINT ID: (B)(4).

Description of Event or Problem

IT WAS REPORTED THAT THE C-MAC MONITOR FREEZES. NO NEGATIVE IMPACT ON STATE OF HEALTH REPORTED.

Additional Manufacturer Narrative

RESULT OF INVESTIGATION: AS THE CLAIMED PRODUCT WAS NOT RETURNED FOR INSPECTION, A MORE DETAILED INVESTIGATION IS NOT POSSIBLE. THE DEVICE PASSED THE QUALITY TEST AT THE TIME OF DELIVERY. BASED ON THE CUSTOMER'S DESCRIPTION OF THE FAULT (MONITOR FREEZES), WE ASSUME THAT IT WAS CAUSED BY USER ERROR, AS THE IFU STATES THAT THE PRODUCT MUST BE CHECKED FOR FUNCTIONALITY BEFORE USE, WHICH WOULD HAVE PREVENTED THIS FAULT. THE EVENT IS FILED UNDER INTERNAL KARL STORZ COMPLAINT ID: (B)(4).