inforMED
MalfunctionCCW

MCGRATH MAC

Received Jul 8, 2025 · Event occurred Jun 10, 2025

Report 2936999-2025-00369 · MDR key 22437757

Device

Generic name

Laryngoscope, Rigid

Model number

301-000-000

Catalog number

301-000-000

Product problems

  • No Display/Image
  • No Display/Image

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem

IT WAS REPORTED THAT PRIOR TO USAGE, THE VIDEO LARYNGOSCOPE'S SCREEN HAD NO DISPLAY, BUT LED LIGHT WAS ON. THERE WAS NO PATIENT INVOLVED.

Additional Manufacturer Narrative

ADDITIONAL INFORMATION: D9, G3, H3, H6 H3 EVALUATION SUMMARY: MEDTRONIC CONDUCTED AN INVESTIGATION BASED UPON ALL INFORMATION RECEIVED. THE DEVICE WAS AVAILABLE FOR EVALUATION. VISUAL INSPECTION NOTED NO NOTABLE CONDITIONS. THE DEVICE WAS TESTED ON AN ONE-WIRE FIXTURE SYSTEM WITH EXTERNAL POWER SUPPLY. THE DEVICE HISTORY LOGS WERE PULLED AND THE LOGS DISPLAYED ACCEPTABLE RESULTS. THE DEVICE WAS POWERED ON BY PUSHING THE POWER BUTTON. THE DEVICE'S CAMERA LED LIGHT OUTPUT WAS FUNCTIONAL, BUT THE SCREEN FAILED TO DISPLAY AN IMAGE AND REMAINED BLANK. IT WAS REPORTED THAT PRIOR TO USAGE, THE VIDEO LARYNGOSCOPE'S SCREEN HAD NO DISPLAY, BUT LED LIGHT WAS ON. THE REPORTED ISSUE WAS CONFIRMED. THE MOST LIKELY CAUSE WAS TRACED TO A COMPONENT FAILURE. THE MANUFACTURING RECORDS FOR EACH DEVICE ARE THOROUGHLY REVIEWED PRIOR TO RELEASE TO ENSURE THAT IT MEETS ALL MEDTRONIC QUALITY SPECIFICATIONS. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.