SONOPLEX® STIM
Received Jul 8, 2025 · Event occurred Aug 2, 2023
Report 9611612-2025-00029 · MDR key 22436774
Device
Generic name
Sonoplex® Stim
Manufacturer
Pajunk Gmbh MedizintechnologieCatalog number
001185-74
Lot number
1472.10
Product problems
- Break
- Break
Patient
NA · Unknown
- Foreign Body In Patient
- Foreign Body In Patient
Narrative
Additional Manufacturer Narrative
THIS CASE IS REPORTED IN RETROSPECT BASED ON A RE-EVALUATION CONDUCTED IN 2025 FOR FORMAL REASONS. THERE WAS NO PARTICULAR RISK TO PATIENTS, USERS, OR THIRD PARTIES.
Description of Event or Problem
IRN# (B)(4). INCIDENT OCCURRED IN LITHUANIA. A SUPRACLAVICULAR BLOCK OF THE BRACHIAL PLEXUS WAS PERFORMED UNDER ULTRASOUND GUIDANCE TO ANESTHETIZE THE PATIENT. AFTER SUCCESSFULLY VISUALIZING THE PLEXUS AND EXTRACTING POTENTIALLY DEFECTIVE REGIONAL ANESTHESIA NEEDLE WITH THE LOCAL ANESTHETIC MIXTURE, THE LATTER DISLODGED FROM THE PLASTIC HUB RETAINER AND REMAINED UNDER THE SKIN. THE INTERVENTION SITE WAS EXAMINED BY ULTRASOUND REVEALING A SOLID NEEDLE STILL PRESENT IN THE SUBCUTANEOUS TISSUE, BUT THE TIP OF THE NEEDLE COULDN'T BE REACHED FROM THE OUTSIDE. THE PLASTIC SURGEON ON DUTY WAS URGENTLY INFORMED AND ARRIVED PROMPTLY. UNDER LIDOCAINE-ANESTHETIZED ULTRASOUND GUIDANCE, THE SHOULDER AREA TISSUES WERE SEPARATED AND THE FOREIGN BODY (NEEDLE) REMOVED. AFTER REMOVAL IT WAS CONFIRMED THAT THE NEEDLE WAS SOLID, AND ULTRASOUND FURTHER CONFIRMED THAT NO PARTS OF THE NEEDLE REMAINED IN THE TISSUES.