BIO-MEDICUS LIFE SUPPORT TM
Received Jul 4, 2025 · Event occurred Jun 12, 2025
Report 9612164-2025-03291 · MDR key 22402533
Device
Generic name
Single Lumen Ecmo Cannula
Manufacturer
Medtronic Mexico S. De R.l. De CvModel number
LS96555-019Catalog number
LS96555-019
Lot number
226108281
Product problems
- Crack
- Crack
Patient
37 YR · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
Description of Event or Problem
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A BIO-MEDICUS LIFE SUPPORT CATHETER AND INTRODUCER, IT WAS REPORTED THAT DURING THE PRIMING PROCESS, IT WAS OBSERVED THAT THE TUBING WAS RUPTURED/CRACKED. THE DEVICE WAS BEING USED TO IMPROVE ARDS (ACUTE RESPIRATORY DISTRESS SYNDROME). THE CUSTOMER PERFORMED AN EXTRACORPOREAL MEMBRANE OXYGENATION PROCEDURE ON THE PATIENT. THE DEVICE WAS REPLACED TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT.
Additional Manufacturer Narrative
MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT WHEN THE DOCTOR OPENED THE CANNULA PACKAGING, THEY OBSERVED A CRACK IN THE CANNULA BODY. THE CANNULA WAS NOT HEATED OR COOLED PRIOR TO USE. CORRECTION D.3 MFR NAME AND POSTAL CODE MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
Additional Manufacturer Narrative
ADDITIONAL INFORMATION B5: MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT WHEN THE DOCTOR OPENED THE CANNULA PACKAGING, THEY OBSERVED A CRACK IN THE CANNULA BODY, SPECIFICALLY THE CONNECTOR. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE EVENT THAT HAD BEEN REPORTED IS A DUPLICATE OF THE EVENT THAT WAS REPORTED IN 9617601-2025-00662. THESE EVENTS HAVE BEEN MERGED. DEVICE EVALUATION SUMMARY: VISUAL INSPECTION SHOWED EVIDENCE OF DAMAGE/CRACK IN THE CONNECTOR. THE REASON FOR THE RETURN WAS CONFIRMED. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.