SHILEY
Received Jul 3, 2025 · Event occurred Jun 27, 2025
Report 8020889-2025-00317 · MDR key 22394856
Device
Generic name
Tube, Bronchial (w/wo Connector)
Manufacturer
Mallinckrodt MedicalModel number
125237Catalog number
125237
Lot number
202206167X
Product problems
- Use of Device Problem
- Use of Device Problem
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
ADDITIONAL INFORMATION: D9, G3, H3, H6 H3 EVALUATION SUMMARY: MEDTRONIC CONDUCTED AN INVESTIGATION BASED UPON ALL INFORMATION RECEIVED. THE DEVICE WAS AVAILABLE FOR EVALUATION. VISUAL INSPECTION NOTED A SLIT ON THE TRACHEAL CUFF WHICH MEASURES 0.26MM IN LENGTH. AN ATTEMPT MADE TO INFLATE THE RETURNED UNIT AS PER THE PARAMETERS SET OUT FAILED AS THE TRACHEAL CUFF COULD NOT BE INFLATED. IT WAS REPORTED THAT PRIOR TO USAGE, THREE DOUBLE LUMEN TUBE'S CUFF HAD AN INFLATION OR DEFLATION ISSUES. THE REPORTED ISSUE WAS CONFIRMED. THE PRODUCT ANALYSIS NOTED EVIDENCE THAT THE DEVICE WAS NOT USED AS INTENDED. THE MANUFACTURING RECORDS FOR EACH DEVICE ARE THOROUGHLY REVIEWED PRIOR TO RELEASE TO ENSURE THAT IT MEETS ALL MEDTRONIC QUALITY SPECIFICATIONS. THE INSTRUCTIONS INCLUDED WITH THIS DEVICE PROVIDE THE FOLLOWING GUIDANCE: VARIOUS BONY ANATOMICAL STRUCTURES (E.G., TEETH) WITHIN THE INTUBATION ROUTE OR ANY INTUBATION TOOLS WITH SHARP SURFACES PRESENT A THREAT TO MAINTAINING CUFF INTEGRITY. CARE MUST BE TAKEN TO AVOID DAMAGING THE THIN-WALLED CUFFS DURING INTUBATION WHICH WOULD CREATE THE NEED TO SUBJECT THE PATIENT TO THE TRAUMA OF EXTUBATION AND RE-INTUBATION. IF EITHER CUFF IS DAMAGED, THE TUBE SHOULD NOT BE USED. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
Additional Manufacturer Narrative
D10 CONCOMITANT PRODUCT: 125235 35FR LT B CATH W HOOK PU CUFFX1 LOT: 202203285X 125235 35FR LT B CATH W HOOK PU CUFFX1 LOT: 202203285X MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
Description of Event or Problem
IT WAS REPORTED THAT PRIOR TO USAGE, THREE DOUBLE LUMEN TUBE'S CUFF HAD AN INFLATION OR DEFLATION ISSUES. THERE WAS NO PATIENT INVOLVED.