inforMED
MalfunctionLCP

TINA-QUANT HEMOGLOBIN A1C GEN.3 - HEMOLYSATE AND WHOLE BLOOD APPLICATION

Received Jul 3, 2025 · Event occurred Jun 17, 2025

Report 1823260-2025-02040 · MDR key 22393792

Device

Generic name

Assay, Glycosylated Hemoglobin

Manufacturer

Roche Diagnostics

Catalog number

05336163190

Lot number

774553

Product problems

  • Appropriate Device Problem Term/Code Not Available
  • Appropriate Device Problem Term/Code Not Available

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE COBAS 4000 C311 STAND ALONE SYSTEM SERIAL NUMBER WAS (B)(6). THE INVESTIGATION IS ONGOING.

Description of Event or Problem

THERE WAS AN ALLEGATION OF QUESTIONABLE TINA-QUANT HEMOGLOBIN A1C GEN.3 - HEMOLYSATE AND WHOLE BLOOD APPLICATION RESULTS FROM THE COBAS 4000 C311 STAND-ALONE SYSTEM. THE INITIAL RESULT WAS 6.2 % AND THE REPEAT RESULT FROM A BIO-RAD D10 ANALYZER WAS 10.8%.

Additional Manufacturer Narrative

DUE TO THE LIMITED INFORMATION PROVIDED, THE INVESTIGATION WAS UNABLE TO DETERMINE A SPECIFIC ROOT CAUSE. THERE WAS NO PRODUCT PROBLEM IDENTIFIED.