PATIENT CABLE

Received Jul 3, 2025 · Event occurred Jun 6, 2025

Report 2182208-2025-03204 · MDR key 22392698

Device

Generic name

Cable, Transducer And Electrode, Patient, (inc

Manufacturer

Medtronic, Inc.

Model number

5492AL

Catalog number

5492AL

Lot number

2209197

Product problems

  • Connection Problem
  • Material Integrity Problem
  • Connection Problem
  • Material Integrity Problem

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

PRODUCT ANALYSIS: ANALYSIS WAS UNABLE TO CONFIRM THE CUSTOMER'S COMMENT THAT THE PATIENT CABLE BECAME DISCONNECTED FROM THE COMPETITOR¿S LEAD WIRE CONNECTOR DURING USE. THE CABLE PASSED ALL VISUAL INCOMING INSPECTION WITH NO ANOMALIES FOUND. ALL THE CONTINUITY TESTS WERE PASSED WITH NO INTERMITTENT OR SHORTED CONNECTIONS. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative

MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem

IT WAS REPORTED THAT THE PATIENT CABLE BECAME DISCONNECTED FROM THE COMPETITOR¿S LEAD WIRE CONNECTOR DURING USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.