GEM MICROVASCULAR ANASTOMOTIC COUPLER
Received Jul 2, 2025 · Event occurred May 8, 2025
Report 1416980-2025-03615 · MDR key 22390259
Device
Generic name
Device,anastomotic,microvascular
Manufacturer
Baxter Healthcare CorporationModel number
GEM2753Catalog number
511100250060
Lot number
SP24H05-2116947
Product problems
- Detachment of Device or Device Component
- Detachment of Device or Device Component
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
E1: INITIAL REPORTER POSTAL CODE: (B)(6). SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem
IT WAS REPORTED THAT THE RINGS OF A 2.5MM COUPLER "CAME OFF". THIS WAS OBSERVED AFTER THE DEVICE WAS TIGHTENED AND ANASTOMOSIS HAD BEEN COMPLETED. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Additional Manufacturer Narrative
ADDITIONAL INFORMATION WAS ADDED TO D4, D9, H3, H4, H6, AND H11: H11: THE DEVICE WAS RECEIVED FOR EVALUATION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. FUNCTIONAL TESTING WAS PERFORMED, AND THE LEFT RING BEING DISLODGED FROM THE JAW ASSEMBLY WAS OBSERVED. THE REPORTED CONDITION WAS VERIFIED. THE CAUSE OF THE CONDITION COULD NOT BE DETERMINED. A BATCH REVIEW WAS CONDUCTED AND THERE WERE NO DEVIATIONS FOUND RELATED TO THIS REPORTED CONDITION DURING THE MANUFACTURE OF THIS LOT. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.