COMPELLA BARIATRIC BED
Received Jul 1, 2025 · Event occurred Jun 12, 2025
Report 1824206-2025-01163 · MDR key 22374053
Device
Generic name
Bed, Flotation Therapy, Powered
Manufacturer
Baxter Healthcare CorporationModel number
P7800ACatalog number
P7800A000001
Product problems
- Activation Problem
- Activation Problem
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
IT WAS REPORTED THAT COMPELLA CAP US SCALE PD PPM (PRODUCT CODE P7800A000001, SERIAL NUMBER (B)(6), HAD THE CPR HANDLE IS DAMAGED UNABLE TO ACTIVATE CPR FEATURE. THIS OCCURRED BEFORE USE. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.
Additional Manufacturer Narrative
THE BAXTER TECHNICIAN FOUND THE CPR HANDLE NEEDED TO BE REPLACED. PER THE BAXTER SERVICE MANUAL THE COMPELLA BED REQUIRES AN EFFECTIVE MAINTENANCE PROGRAM. WE RECOMMEND THAT YOU PERFORM ANNUAL PREVENTATIVE MAINTENANCE. PAY PARTICULAR ATTENTION TO SAFETY FEATURES, WHICH INCLUDE BUT ARE NOT LIMITED TO: ALL CAREGIVER AND PATIENT CONTROLS, ALONG WITH THEIR INDICATORS, OPERATE CORRECTLY; THIS INCLUDES CONTROLS FOR CPR. REPLACE OR REPAIR PARTS AS NECESSARY. A SEARCH OF THE BAXTER MAINTENANCE RECORDS SHOWED BAXTER PERFORMED PREVENTATIVE MAINTENANCE ON THIS BED IN MAY 6, 2025. IT IS UNKNOWN IF THE FACILITY PERFORMED ANY OTHER PREVENTATIVE MAINTENANCE ON THIS BED. THE TECHNICIAN WILL REPLACED THE CPR HANDLE TO RESOLVE THE REPORTED EVENT. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.