inforMED
MalfunctionFEM

COUPLER AC FFL 19

Received Jun 27, 2025 · Event occurred Jan 23, 2025

Report 1221934-2025-02355 · MDR key 22344613

Device

Generic name

Accessories, Photographic, For Endoscope (exclude Light Sources)

Catalog number

242437

Product problems

  • Device Reprocessing Problem
  • Moisture Damage
  • Device Reprocessing Problem
  • Moisture Damage

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

IT WAS REPORTED THAT THE COUPLER AC FFL 19 DEVICE HAD WATER DROPLETS INSIDE THE COUPLERS. DURING AN IN-HOUSE ENGINEERING EVALUATION IT WAS FOUND THAT THE DEVICE HAD FOREIGN SUBSTANCE/DEBRIS/CLEANING/STERILIZATION. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

Additional Manufacturer Narrative

THIS REPORT IS BEING SUBMITTED IN PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY MITEK OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, MITEK, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. H11 ADDITIONAL NARRATIVE: INVESTIGATION SUMMARY: THE COMPLAINT DEVICE WAS RECEIVED AT THE SERVICE CENTER AND EVALUATED. DURING THE SERVICE EVALUATION THE FOLLOWING DEFECTS WERE IDENTIFIED: MOISTURE INSIDE OPTICAL SYSTEM, OBJECTIVE IS NOT REPAIRABLE. FUNCTIONAL: MOISTURE/WATER DAMAGE AND VISUAL: FOREIGN SUBSTANCE/DEBRIS/CLEANING/STERILIZATION. PER SERVICE REPORTS, THIS COMPLAINT WAS CONFIRMED. BASED ON SERVICE MANUAL OPERATIONAL AND DIAGNOSTIC, THE DEVICE WAS FOUND TO BE NON-REPAIRABLE, AND IT WAS SCRAPPING. BASED ON THE INVESTIGATION FINDINGS, THE POTENTIAL ROOT CAUSE IS TRACED TO IMPROPER MAINTENANCE. IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED AND THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE REPORTED COMPLAINT CONDITION. AS PART OF J&J MEDTECH ORTHOPAEDICS QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES.