SINGLE PRONG SOFT TISSUE RET
Received Jun 26, 2025 · Event occurred Jun 9, 2025
Report 1818910-2025-10700 · MDR key 22338722
Device
Product problems
- Break
- Break
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
IT WAS REPORTED THAT THE TIP OF THE DEVICE IS BROKEN. THERE WAS NO PATIENT CONSEQUENCE.
Additional Manufacturer Narrative
PRODUCT COMPLAINT #(B)(4). THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803 (AND/OR PART 4, AS APPLICABLE). THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
Additional Manufacturer Narrative
PRODUCT COMPLAINT #(B)(4). INVESTIGATION SUMMARY BROKEN TIP THE PRODUCT WAS RETURNED TO J&J MEDTECH ORTHOPAEDICS FOR EVALUATION. VISUAL ANALYSIS OF THE RETURNED DEVICE REVEALED THAT THE SINGLE PRONG SOFT TISSUE RET HAD THE TIP PARTIALLY BROKEN. FRAGMENT WAS NOT RETURNED FOR EVALUATION. ADDITIONALLY, DEVICE DISPLAYS SIGNS OF REPEATED AND EXTENSIVE USE. THE BREAKAGE APPEARS TO BE CONSISTENT WITH THE EFFECTS OF REPEATED USE AND SERVICING. THE LIFECYCLE REQUIREMENTS OF THE DEVICE ARE EVENT RELATED AND DEPEND ON THE USE AND INSPECTION OF THE DEVICE IN CLINICAL PRACTICE. AS THE DEVICE CAN BE DAMAGED ON THE FIRST OR 100TH USE, THE DEVICE MUST BE PROPERLY INSPECTED PRIOR TO EACH SURGICAL USE. REFER TO THE DEVICE/COUNTRY SPECIFIC IFU FOR INFORMATION RELATED TO END OF LIFE, REPROCESSING INSTRUCTIONS, AND INSPECTION PROCEDURES. A FUNCTIONAL TEST WAS NOT PERFORMED SINCE IT WAS NOT APPLICABLE TO THE COMPLAINT CONDITION. A DIMENSIONAL INSPECTION WAS NOT PERFORMED SINCE IT WAS NOT APPLICABLE TO THE COMPLAINT CONDITION. THE OVERALL COMPLAINT WAS CONFIRMED AS THE OBSERVED CONDITION OF THE SINGLE PRONG SOFT TISSUE RET WOULD HAVE CONTRIBUTED TO THE COMPLAINED ISSUE. BASED ON THE INVESTIGATION FINDINGS, THE POTENTIAL CAUSE IS TRACED TO END OF LIFE, AND IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE REPORTED COMPLAINT CONDITION. AS PART OF J&J MEDTECH ORTHOPAEDICS QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. DEVICE HISTORY LOT THE PRODUCT INVESTIGATION FOUND NO EVIDENCE SUSPECTING AN ERROR IN THE MANUFACTURING OR MATERIAL THAT WOULD BE A CONTRIBUTING FACTOR IN THE REPORTED ALLEGATION(S). A MANUFACTURING RECORDS EVALUATION (MRE) WAS NOT PERFORMED.