inforMED
InjuryPKC

STEMLESS L 40

Received Jun 26, 2025 · Event occurred Jun 6, 2025

Report 1818910-2025-10647 · MDR key 22337110

Device

Generic name

Shoulder Implant - Humeral Anchor

Catalog number

520000040

Lot number

662039

Product problems

  • Osseointegration Problem
  • Osseointegration Problem

Patient

68 YR · Male

  • Bone Fracture(s)
  • Inadequate Osseointegration
  • Bone Fracture(s)
  • Inadequate Osseointegration

Narrative

Additional Manufacturer Narrative

PRODUCT COMPLAINT #(B)(4). THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803 (AND/OR PART 4, AS APPLICABLE). THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.

Description of Event or Problem

IT WAS REPORTED THAT THE STEMLESS REVERSE SPUN OUT, THE SUPERIOR MEDIAL WALL OF THE HUMERUS WAS BLOWN OUT. THERE WAS LOOSENING OF STEMLESS COMPONENT. SURGEON STATE IT LOOKED TRAUMATIC BUT PATIENT REPORTED NO INCIDENT. AFFECTED SIDE: LEFT SHOULDER.

Additional Manufacturer Narrative

PRODUCT COMPLAINT #(B)(4). INVESTIGATION SUMMARY: NO DEVICE ASSOCIATED WITH THIS REPORT WAS RECEIVED FOR EXAMINATION. THE PRODUCT INVESTIGATION FOUND NO EVIDENCE SUSPECTING AN ERROR IN THE MANUFACTURING OR MATERIAL THAT WOULD BE A CONTRIBUTING FACTOR IN THE REPORTED ALLEGATION(S). A RECORDS EVALUATION (MRE) WAS NOT PERFORM AS PART OF OUR COMPANY QUALITY SYSTEM PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND DISTRIBUTED TO APPROVED SPECIFICATIONS. ADDITIONAL COMPLAINT INFORMATION MONITORING FOR POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AS PART OF THE POST-MARKET SURVEILLANCE. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. DEVICE HISTORY LOT: THE PRODUCT INVESTIGATION FOUND NO EVIDENCE SUSPECTING AN ERROR IN THE MANUFACTURING OR MATERIAL THAT WOULD BE A CONTRIBUTING FACTOR IN THE REPORTED ALLEGATION(S). A MANUFACTURING RECORDS EVALUATION (MRE) WAS NOT PERFORMED. H11 ADDITIONAL NARRATIVE: ADDED: D10.

Additional Manufacturer Narrative

PRODUCT COMPLAINT # (B)(4). THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION THAT HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES JOINT RECONSTRUCTION, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES JOINT RECONSTRUCTION, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. H11 ADDITIONAL NARRATIVE: ADDED :D9. IF INFORMATION THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT IS OBTAINED, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative

PRODUCT COMPLAINT # (B)(4). INVESTIGATION SUMMARY- ACCORDING TO THE INFORMATION RECEIVED, "IT WAS REPORTED THAT THE STEMLESS REVERSE SPUN OUT, THE SUPERIOR MEDIAL WALL OF THE HUMERUS WAS BLOWN OUT. THERE WAS LOOSENING OF STEMLESS COMPONENT. SURGEON STATE IT LOOKED TRAUMATIC BUT PATIENT REPORTED NO INCIDENT." THE PRODUCT WAS RETURNED TO J&J MEDTECH ORTHOPAEDICS FOR EVALUATION. VISUAL INSPECTION OF THE RETURNED DEVICE FOUND THAT THE STEMLESS L 40 SHOWS SIGNS OF BONE INGROWTH ADHERED TO THE FIXATION SURFACE WITH A GOOD FIXATION OF THE IMPLANT. THEREFORE, THE REPORTED IMPLANT LOOSENING CONDITION WAS UNABLE TO BE CONFIRMED. ADDITIONALLY, THE DEVICE ARRIVED ASSEMBLED WITH THE REVERSE LINER AND THE REVERSE HUMERAL SHELL. A FUNCTIONAL TEST WAS NOT PERFORMED SINCE IT WAS NOT APPLICABLE TO THE COMPLAINT CONDITION. A DIMENSIONAL INSPECTION WAS NOT PERFORMED SINCE IT WAS NOT APPLICABLE TO THE COMPLAINT CONDITION. THE OVERALL COMPLAINT WAS NOT CONFIRMED AS THE OBSERVED CONDITION OF THE STEMLESS L 40 WOULD HAVE NOT CONTRIBUTED TO THE COMPLAINED DEVICE ISSUE. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE REPORTED COMPLAINT CONDITION. AS PART OF J&J MEDTECH ORTHOPAEDICS QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. DEVICE HISTORY LOT- A SEARCH OF THE MEDTECH ORTHOPAEDICS (NC) QUALITY SYSTEM FOUND NO NC¿S ASSOCIATED WITH THIS PRODUCT/LOT COMBINATION. H11 ADDITIONAL NARRATIVE: CORRECTED: H3.