inforMED
MalfunctionMBB

SMARTSET GHV GENTAMICIN 40G

Received Jun 25, 2025 · Event occurred Jun 4, 2025

Report 1818910-2025-10396 · MDR key 22330860

Device

Generic name

Bone Cement : Bone Cement

Catalog number

3095040

Lot number

4419580

Product problems

  • Device Damaged Prior to Use
  • Device Damaged Prior to Use

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

PRODUCT COMPLAINT # (B)(4). THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803 (AND/OR PART 4, AS APPLICABLE). THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE. H11 ADDITIONAL NARRATIVE: D4: UDI: AS THE LOT NUMBER FOR THE DEVICE INVOLVED IN THE EVENT PROVIDED 4419580 WAS NOT ABLE TO BE VALIDATED, THE FULL UDI IS CURRENTLY NOT AVAILABLE. G4: DEVICE IS NOT DISTRIBUTED IN THE UNITED STATES BUT IS SIMILAR TO DEVICE MARKETED IN THE USA. DMF# - 13704 TRADE NAME ¿ GENTAMICIN SULPHATE ACTIVE INGREDIENT(S) ¿ GENTAMICIN SULPHATE DOSAGE FORM - POWDER STRENGTH ¿ 1.0G ACTIVE IN OUR CEMENTS.

Description of Event or Problem

IT WAS REPORTED THAT WHEN OPENING THE PACKAGING OF BONE CEMENT DURING A KNEE JOINT REPLACEMENT SURGERY, IT WAS FOUND THAT THE INNER POWDER PACKAGING WAS DAMAGED. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. THERE WAS NO SURGICAL DELAY.

Additional Manufacturer Narrative

PRODUCT COMPLAINT # (B)(4). INVESTIGATION SUMMARY: INNER PACKING DAMAGED. OPENING THE PACKAGING OF BONE CEMENT, IT WAS FOUND THAT THE POWDER PACKAGING WAS DAMAGED (AS THE PHOTO SHOWS). ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED. THE PRODUCT WAS NOT RETURNED TO J&J MEDTECH ORTHOPAEDICS; HOWEVER, PHOTOS WERE PROVIDED FOR REVIEW. THE PHOTO INVESTIGATION REVEALED THAT IT APPEARS THAT THE POWDER PACK HAD AN ISSUE WITH THE SEAL, THE POWDER HAD CAUSED THE SEAL TO BURST AND SEAL WAS COMPROMISED WITH POWDER. AS THE DEVICE WAS NOT RETURNED, AN AS-RECEIVED CONDITION COULD NOT BE ASSESSED, AND A DIMENSIONAL INSPECTION AND DOCUMENT/SPECIFICATION REVIEW WERE NOT COMPLETED. THE OVERALL COMPLAINT WAS CONFIRMED AS THE OBSERVED CONDITION OF THE SMARTSET GHV GENTAMICIN 40G WOULD HAVE CONTRIBUTED TO THE COMPLAINED DEVICE ISSUE. BASED ON THE INFORMATION CURRENTLY AVAILABLE, A PRODUCT ISSUE WAS IDENTIFIED DURING THE INVESTIGATION. THIS PRODUCT ISSUE WILL BE ADDRESSED THROUGH J&J MEDTECH ORTHOPAEDICS QUALITY SYSTEM. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. DEVICE HISTORY LOT: NR HAS BEEN RAISED TO ADDRESS CURRENT COMPLAINT CONDITION. ADDED: D4 EXPIRATION DATE, H4. CORRECTED INFORMATION: D4 (PRIMARY UDI #).