inforMED
MalfunctionFWY

OPERON

Received Jul 28, 2011 · Event occurred Jun 24, 2011

Report 8010153-2011-00001 · MDR key 2232940

Device

Generic name

Surgical Table

Model number

D850

Catalog number

3260.065

Product problems

  • Component Falling

Patient

Not reported

  • No Known Impact Or Consequence To Patient

Narrative

Description of Event or Problem

DURING A SURGICAL PROCEDURE, WHEN THE TRENDELENBURG FUNCTION BUTTON WAS PRESSED, THE TABLE TOP WAS REPORTED TO HAVE DROPPED STRAIGHT DOWN BETWEEN 6" - 14". THERE WAS NO PT INJURY REPORTED, BUT THE PROCEDURE WAS ABORTED AFTER THE INCIDENT.

Additional Manufacturer Narrative

THE TABLE WAS EVALUATED BY A BERCHTOLD CORP. SERVICE TECHNICIAN ON (B)(4) 2011 AFTER THE INCIDENT AT THE LOCATION WHERE THE INCIDENT OCCURRED. THE TABLE IS IN GOOD PHYSICAL CONDITION. ALL THE TABLE FUNCTIONS OPERATED AS EXPECTED WITHOUT PROBLEMS, AND THE REPORTED PROBLEM COULD NOT BE DUPLICATED. THE TABLE HAS BEEN RETURNED TO THE BERCHTOLD CORPORATION FACILITY IN (B)(4) FOR FURTHER EVAL.