PORTEX TUBES BLUSELECT SUCTIONAID
Received Jun 24, 2025 · Event occurred Jun 3, 2025
Report 3012307300-2025-07537 · MDR key 22318597
Device
Generic name
Bag, Reservoir
Manufacturer
Icu Medical, Inc.Model number
101/870/070Catalog number
101/870/070CZ
Lot number
6004213
Product problems
- Crack
- Crack
Patient
NA · Unknown
- Salivary Hypersecretion
- Salivary Hypersecretion
Narrative
Additional Manufacturer Narrative
D9: DATE RETURNED TO MFG.: 6/26/2025. H3 AND H6. CODES: UPDATED. DEVICE EVALUATION: THE SUSPECTED DEVICE WAS RECEIVED IN USED CONDITION. UNDER VISUAL INSPECTION THE SUCTION CONNECTOR SEEMED TO BE WITHOUT ANY DEFECT. WHEN A LUER SLIP SYRINGE WAS INSERTED INTO THE CONNECTOR A CRACK WAS OBSERVED. THE COMPLAINT WAS CONFIRMED. THIS DEFECT CAN BE CAUSED BY APPLYING EXCESSIVE FORCE WHEN CONNECTING THE SYRINGE OR DURING THE CONNECTION OF THE SUCTION DEVICE, THE ROOT CAUSE COULD NOT BE ASSOCIATED WITH THE MANUFACTURING PROCESS. A DEVICE HISTORY RECORD (DHR) REVIEW SHOWED NO DISCREPANCIES OR NON-CONFORMANCES DURING THE MANUFACTURING OF THE REPORTED LOT NUMBER.
Description of Event or Problem
IT WAS REPORTED THAT THE DEVICE INSERTED ON (B)(6) 2025 WAS FOUND TO BE FAULTY. THE TUBE CRACKED WHEN A 10ML SYRINGE WAS INSERTED, AN ISSUE THAT OCCURRED MULTIPLE TIMES RECENTLY. THIS REQUIRED AN URGENT TUBE CHANGE AND THE USE OF DEVICE TO PREVENT INFECTION FOR THE PATIENT. A FAULTY SUBGLOTTIC PORT WAS IDENTIFIED DURING AN OUTPATIENT VISIT, AND THE PATIENT SHOWED INCREASED SECRETIONS AND RISK OF ASPIRATION. A TRACHEOSTOMY TUBE CHANGE WAS CARRIED OUT TO ENSURE SAFE AIRWAY MANAGEMENT. THE ISSUE WAS NOW RESOLVED WITH A NEW TRACHEOSTOMY TUBE.
Additional Manufacturer Narrative
E1 - INITIAL REPORTER PHONE: (B)(6). H3, H6: INVESTIGATION INCLUDING ROOT CAUSE ANALYSIS IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.