inforMED
MalfunctionCCW

MCGRATH MAC

Received Jun 19, 2025 · Event occurred Jun 4, 2025

Report 3010244187-2025-00565 · MDR key 22293530

Device

Generic name

Laryngoscope, Rigid

Model number

301-000-000

Catalog number

301-000-000

Product problems

  • No Display/Image
  • No Display/Image

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem

IT WAS REPORTED THAT DURING TESTING, THE VIDEOLARYNGOSCOPE POWERED UP BUT DID NOT DISPLAY ANY INFORMATION. VARIOUS BATTERIES WERE ATTEMPTED WITH NO CHANGE IN THE ISSUE. THERE WAS NO PATIENT INVOLVED.

Additional Manufacturer Narrative

ADDITIONAL INFORMATION: D9, G3, H3 AND H6. H3 EVALUATION SUMMARY: MEDTRONIC CONDUCTED AN INVESTIGATION BASED UPON ALL INFORMATION RECEIVED. THE DEVICE WAS AVAILABLE FOR EVALUATION. THE DEVICE'S CAMERA LED LIGHT OUTPUT WAS FUNCTIONAL BUT THE LCD SCREEN WAS NOT ABLE TO DISPLAY AN IMAGE. THE SCREEN WAS BLANK. IT WAS REPORTED THAT THE VIDEOLARYNGOSCOPE POWERED UP BUT DID NOT DISPLAY ANY INFORMATION. VARIOUS BATTERIES WERE ATTEMPTED WITH NO CHANGE IN THE ISSUE. THE REPORTED ISSUE WAS CONFIRMED. THE MOST LIKELY CAUSE WAS COMPONENT FAILURE. THE MANUFACTURING RECORDS FOR EACH DEVICE ARE THOROUGHLY REVIEWED PRIOR TO RELEASE TO ENSURE THAT IT MEETS ALL MEDTRONIC QUALITY SPECIFICATIONS. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.