MalfunctionJPK

4C PLUS COULTER CELL CONTROL

Received Aug 12, 2011 · Event occurred Mar 14, 2010

Report 1061932-2011-01133 · MDR key 2229280

Device

Model number

NA

Catalog number

7547114

Lot number

1000302K

Product problems

  • Fluid/Blood Leak
  • Loose or Intermittent Connection

Patient

UNK

  • No Known Impact Or Consequence To Patient

Narrative

Additional Manufacturer Narrative

TORQUE TEST RECORDS FOR THE VIAL LOTS IN THIS KIT DEMONSTRATED PASSING RESULTS. A REVIEW FOR SIMILAR EVENTS WAS CONDUCTED. THERE HAVE BEEN FOUR REPORTS OF LEAKING FOR THIS PRODUCT FROM 2005 TO 2010. ROOT CAUSE WAS NOT DETERMINED. THIS REPORTABLE EVENT WAS IDENTIFIED DURING A RETROSPECTIVE REVIEW CONDUCTED BETWEEN 01/01/2008 AND 10/23/2010 OF COMPLAINTS FOR ADDITIONAL REPORTABLE EVENTS.

Description of Event or Problem

CUSTOMER REPORTED A POTENTIAL BIOHAZARD WHEN THE CONTROL VIALS OF A 4C PLUS COULTER CELL CONTROL KIT WERE FOUND TO BE LEAKING. A SMALL AMOUNT OF THE CONTROL LEAKING ONTO THE VIAL CAPS WHILE STILL INSIDE THE CONTAINER. THIS APPEARED TO BE DUE TO LOOSE VIAL CAPS. THE OPERATOR WAS WEARING APPROPRIATE PERSONAL PROTECTIVE EQUIPMENT OF GLOVES AND SAFETY GLASSES. THERE WAS NO EXPOSURE OF THE MATERIAL TO OPEN WOUNDS OR MUCOUS MEMBRANES. NO REPORTS OF DEATH OR SERIOUS INJURY, AND NO AFFECT TO OPERATOR SAFETY AS A RESULT OF THIS EVENT.