inforMED
InjuryDZE

VEGA® RV IMPLANT Ø4.0 X 10MM

Received Jun 18, 2025 · Event occurred May 22, 2019

Report 3003267134-2025-00009 · MDR key 22284584

Device

Generic name

Dental Implant

Manufacturer

Soadco S.l.

Model number

18 40 10

Lot number

26526-1122

Product problems

  • Fracture
  • Fracture

Patient

52 YR · Unknown

  • Implant Pain
  • Implant Pain

Narrative

Additional Manufacturer Narrative

THE LOSS OF FIT DURING SEATING BETWEEN THE PROSTHESIS AND THE IMPLANT DOES NOT DISTRIBUTE THE MASTICATORY LOAD EVENLY AND, CONSEQUENTLY, INCREASES THE LIKELIHOOD OF GENERATING DEFECTS IN THE IMPLANT THAT WILL LEAD TO ITS FAILURE DUE TO MATERIAL FATIGUE. THEREFORE, THIS LOSS OF FIT HAS A DIRECT IMPACT ON THE DISTRIBUTION OF THE MASTICATORY LOAD ON THE IMPLANT AND, CONSEQUENTLY, ON ITS MECHANICAL AND FATIGUE RESISTANCE.

Description of Event or Problem

AN BROKEN IMPLANT IS RECEIVED AT SOADCO S.L. FACILITIES WHERE THE MANUFACTURER AND THE STATE OF THE PROSTHESIS PLACED IS UNKNOWN (CAUSE: BREAKAGE OF THE IMPLANT IN THE PATIENT). THIS IMPLANT PRESENTS A PLASTIC DEFORMATION OF THE MATERIAL. PREVIOUSLY A CLAIM HAS BEEN RECEIVED FOR THIS SAME IMPLANT MODEL AND MANUFACTURING LOT. THIS INCIDENCE IS DIRECTLY RELATED TO THE LOSS OF ADJUSTMENT BETWEEN THE PROSTHESIS AND THE IMPLANT.

Remedial action

  • Notification