inforMED
InjuryDZE

VEGA NV Ø3.5 X 12MM /VEGA RV Ø4.0 X 10MM IMPLANT

Received Jun 18, 2025 · Event occurred Feb 12, 2019

Report 3003267134-2025-00005 · MDR key 22284570

Device

Generic name

Dental Implant

Manufacturer

Soadco, S.l

Lot number

26772-1124 / 26312-1120

Product problems

  • Fracture
  • Fracture

Patient

NA · Male

  • Implant Pain
  • Implant Pain

Narrative

Additional Manufacturer Narrative

AFTER THE TECHNICAL ANALYSIS OF THE INCIDENT, IT IS CONCLUDED THAT:THE LOSS OF FIT IN THE SEATING BETWEEN THE PROSTHESIS AND THE IMPLANTS DOES NOT DISTRIBUTE THE MASTICATORY LOAD EVENLY AND, CONSEQUENTLY, INCREASES THE LIKELIHOOD OF GENERATING DEFECTS IN THE IMPLANT THAT WILL INDUCE ITS BREAKAGE DUE TO MATERIAL FATIGUE.

Description of Event or Problem

A DISTRIBUTOR INFORMS SOADCO, S.L. THAT A DOCTOR HAS INFORMED THEM OF THE BREAKAGE OF TWO IMPLANTS IN A BRIDGE. IN ADDITION, THE DISTRIBUTOR INFORMS THAT THE MANUFACTURER OF THE PROSTHESIS IS UNKNOWN AND IT IS NOT POSSIBLE TO MAKE A HISTORY OF THE MATERIAL USED, BUT IT IS CONFIRMED THAT IT IS A COBALT CHROME PROSTHESIS.

Remedial action

  • Notification