inforMED
InjuryDZE

VEGA IMPLANT Ø3.5 X 8MM

Received Jun 18, 2025 · Event occurred Sep 27, 2022

Report 3003267134-2025-00002 · MDR key 22284567

Add to watch list
Product code (DZE)
Manufacturer (Soadco, S.l.)
Model (18 35 08)

Device

Generic name

Dental Implant

Manufacturer

Soadco S.l.

Model number

18 35 08

Product problems

  • Loss of Osseointegration
  • Patient Device Interaction Problem
  • Loss of Osseointegration
  • Patient Device Interaction Problem

Patient

NA · Unknown

  • Implant Pain
  • Implant Pain

Narrative

Description of Event or Problem

THE CLIENT SENDS THE FRACTURED IMPLANT WITH ITS PROSTHESIS.THE CLIENT IS REPEATEDLY ASKED TO SUBMIT THE QUALITY REPORT (A DOCUMENT CONTAINING ALL THE PATIENT INFORMATION AND INFORMATION ABOUT THE PRODUCT AND THE INCIDENT) WITHOUT RECEIVING A RESPONSE. THE CLIENT DOES NOT PROVIDE ANY BATCH AND REFERENCE DATA OF THE PRODUCT.

Additional Manufacturer Narrative

ALTHOUGH THE INFORMATION RECEIVED IS LIMITED, IT CAN BE CONCLUDED THAT THE IMPLANT HAS BEEN WORKING UNDER UNFAVORABLE CONDITIONS SINCE THE PROSTHETIC HEIGHT IS WITHIN THE MAXIMUM RECOMMENDED LIMIT (IT SHOULD NOT EXCEED THE LENGTH OF THE IMPLANT) AND THE PROSTHETIC DIAMETER IS TWO AND A HALF TIMES GREATER THAN THE IMPLANT DIAMETER.

Remedial action

  • Notification