MCGRATH MAC
Received Jun 17, 2025 · Event occurred May 27, 2025
Report 3010244187-2025-00562 · MDR key 22231876
Device
Generic name
Laryngoscope, Rigid
Manufacturer
Plexus Electronica S. De R.l. De C.v.Model number
301-000-000Catalog number
301-000-000
Product problems
- No Display/Image
- No Display/Image
Patient
48 YR · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
Description of Event or Problem
IT WAS REPORTED THAT DURING USE OF THE VIDEOLARYNGOSCOPE, THE SCREEN INITIALLY TURNED ON, THEN TURNED BLANK WHITE, AND SUBSEQUENTLY BLACK AS THE BLADE WAS BEING PLACED IN THE MOUTH. THE LIGHT-EMITTING DIODE (LED) LIGHT REMAINED ON. THE INTUBATION ATTEMPT WAS COMPLETED WITH DIRECT LARYNGOSCOPY AS AN ALTERNATE MEDICAL MEASURE. THERE WAS NO HARM TO THE PATIENT AND NO DELAY IN TREATMENT.
Additional Manufacturer Narrative
ADDITIONAL INFORMATION: D9, G3, H3 AND H6. H3 EVALUATION SUMMARY: MEDTRONIC CONDUCTED AN INVESTIGATION BASED UPON ALL INFORMATION RECEIVED. THE DEVICE WAS AVAILABLE FOR EVALUATION. THE EVALUATION FOUND NO POTENTIALLY CONTRIBUTING FACTORS, AND THE SAMPLE MET ALL RELATED SPECIFICATIONS. IT WAS REPORTED THAT DURING USE OF THE VIDEO LARYNGOSCOPE, THE SCREEN INITIALLY TURNED ON, THEN TURNED BLANK WHITE, AND SUBSEQUENTLY BLACK AS THE BLADE WAS BEING PLACED IN THE MOUTH. THE LIGHT-EMITTING DIODE (LED) LIGHT REMAINED ON. THE REPORTED ISSUE COULD NOT BE CONFIRMED. THE MOST LIKELY CAUSE COULD NOT BE IDENTIFIED BECAUSE NO RELATED PROBLEM WAS DETECTED WITH THE DEVICE. THE MANUFACTURING RECORDS FOR EACH DEVICE ARE THOROUGHLY REVIEWED PRIOR TO RELEASE TO ENSURE THAT IT MEETS ALL MEDTRONIC QUALITY SPECIFICATIONS. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.