inforMED
MalfunctionMPB

ARROW 2 LUMEN BLUE HIGH VOLUME WITH BLUE FLEXTIP

Received Jun 16, 2025 · Event occurred Apr 29, 2025

Report 22211364 · MDR key 22211364

Device

Generic name

Catheter, Hemodialysis, Non-implanted

Product problems

  • Device Slipped
  • Detachment of Device or Device Component

Patient

1 MO · Female

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

HD [HEMODIALYSIS] CATHETER WHITE SECUREMENT DEVICE WAS SUTURED TO PATIENT¿S SKIN; CATHETER CAME OUT OF SECUREMENT DEVICE AND OUT OF PATIENT. THE SUTURES WERE PLACED AT THE TRIANGULAR JUNCTURE HUB WITH INTEGRAL ROTATING SUTURE WINGS, AND THE BLUE PLASTIC PIECE WAS ABLE TO SLIP OUT OF THAT. THIS SAME EXACT ISSUE OCCURRED WITH 2 DIFFERENT PATIENT¿S AND THE LINES WERE PLACED BY TWO DIFFERENT PROVIDERS.