inforMED
MalfunctionJDP

INTERLCKNG BOLT 4.5 VA COND INSERTN HNDL

Received Jun 11, 2025 · Event occurred May 17, 2025

Report 8030965-2025-05812 · MDR key 22183929

Device

Generic name

Condylar Plate Fixation Implant

Manufacturer

Synthes Gmbh

Catalog number

03.231.005

Lot number

L463664

Product problems

  • Device-Device Incompatibility
  • Material Twisted/Bent
  • Device-Device Incompatibility
  • Material Twisted/Bent

Patient

77 YR · Female

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

IT WAS REPORTED THAT ON (B)(6) 2025 THE THREADS ON BOLT WERE STRUGGLING TO ENGAGE WITH PLATES PROPERLY. PRESUMABLY DUE TO WEAR OVER YEARS OF USE. WHILE TRYING TO ATTACH BOLT TO DISTAL FEMUR PLATE THE FIT WASN¿T SECURE AND DAMAGED THE THREADS IN THE PLATE. THE PLATE COULD NO LONGER BE USED WITH AN AIMING ARM, SO ANOTHER TRAY NEEDED TO BE OPENED TO COMPLETE THE CASE. THERE WAS NO PATIENT STATUS/ OUTCOME / CONSEQUENCES.

Additional Manufacturer Narrative

PRODUCT COMPLAINT # (B)(4). THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803 (AND/OR PART 4, AS APPLICABLE). THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. H11 ADDITIONAL NARRATIVE: ADDED: D10 AND H4 CORRECTED: G1 H6 COMPONENT CODES: MOST RELEVANT COMPONENT CODE IS G07002 (APPROPRIATE TERM/CODE NOT AVAILABLE) TO CAPTURE NO FINDINGS AVAILABLE DUE TO NO PRODUCT RETURNED. H3, H6: PRODUCT WAS NOT RETURNED. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. DEVICE HISTORY REVIEW PART: 03.231.005-US LOT: L463664 MANUFACTURING SITE: HAGENDORF RELEASE TO WAREHOUSE DATE: 02 AUG 2017. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED ARTICLE LOT AND NO NON-CONFORMANCES WERE IDENTIFIED. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE.