inforMED
MalfunctionGFJ

V. MUELLER

Received Jun 5, 2025 · Event occurred Apr 15, 2025

Report 1527821-2025-00008 · MDR key 22153992

Device

Generic name

Orthopaedic Mallet

Manufacturer

Steris Corporation

Model number

OS1519

Catalog number

OS1519

Lot number

(10)XSSZ09

Product problems

  • Insufficient Device Problem Information
  • Insufficient Device Problem Information

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

THE USER FACILITY REPORTED THAT DURING PATIENT PROCEDURES PIECES OF METAL WERE "FLECKING OFF" OF THEIR ORTHOPAEDIC MALLET. NO REPORT OF INJURY.

Additional Manufacturer Narrative

INVESTIGATION OF THE REPORTED EVENT IS IN PROGRESS. A FOLLOW-UP REPORT WILL BE SUBMITTED ONCE ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative

THE DEVICE SUBJECT OF THE REPORTED EVENT WAS RETURNED TO THE SUPPLIER FOR EVALUATION. UPON EVALUATION, IT WAS FOUND THAT PIECES OF METAL WERE COMING OFF THE DEVICE. THE LIKELY CAUSE OF THE REPORTED EVENT IS ATTRIBUTED TO WEAR AND TEAR OF THE INSTRUMENT OVER TIME. THE DEVICE SUBJECT OF THE REPORTED EVENT WAS MANUFACTURED IN 2016 AND IS APPROXIMATELY NINE YEARS OLD. THE INSTRUMENT WAS PERMANENTLY REMOVED FROM SERVICE. NO ADDITIONAL ISSUES HAVE BEEN REPORTED.