inforMED
MalfunctionFNL

RAYE'S, INC DBA SIZEWISE MANUFACTURING

Received Jun 4, 2025 · Event occurred May 16, 2025

Report 1931307-2025-00014 · MDR key 22146502

Device

Generic name

120v Low Bed 39"

Model number

33060200

Product problems

  • Use of Device Problem
  • Use of Device Problem

Patient

NA · Male

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

DEVICE MISUSE: BEDFRAME USED IN HOME (VS HEALTHCARE) SETTING. USER SAT AT HEAD END OF BED WHILE HOLDING HEADBOARD AND UTILIZED THE BED HEAD-END RAISE FUNCTION VIA HAND CONTROLLER TO LIFT ENTIRE BODY WEIGHT TO ASSIST WITH STANDING. THE IMPROPER WEIGHT DISTRIBUTION RESULTED IN ACTUATOR FAILURE.

Description of Event or Problem

BEDFRAME NOT WORKING PROPERLY WHILE IN PATIENT USE, HOWEVER PATIENT WAS NOT INJURED. BED'S HEAD ARTICULATION HAS MOVEMENT BUT IS JERKY. PATIENT NEEDS TO HELP THE ACTUATOR IN ORDER FOR IT TO MOVE. ACTUATOR GIVES OUT WHEN PUTTING A LOAD OF WEIGHT ON IT.

Remedial action

  • Repair