inforMED
InjuryMAF

GRAFTMASTER

Received Jun 4, 2025 · Event occurred Apr 1, 2025

Report 2024168-2025-05730 · MDR key 22144483

Device

Generic name

Coronary Stent Delivery System

Model number

1012817-26C

Catalog number

1012817-26C

Lot number

4091641

Product problems

  • Failure to Advance
  • Failure to Advance

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

IT WAS REPORTED THAT THE PROCEDURE PERFORMED WAS TO TREAT A DE NOVO, MODERATELY CALCIFIED, MODERATELY TORTUOUS LEFT ANTERIOR DESCENDING CORONARY ARTERY. THE 2.80X26 GRAFTMASTER COVERED STENT FAILED TO CROSS DUE TO ANATOMY. A NEW 2.8X19MM GRAFTMASTER WAS USED TO TREAT THE PERFORATION AND SUCCESSFULLY COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT SEQUELA AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Additional Manufacturer Narrative

MANUFACTURER'S INVESTIGATION IS STILL PENDING AT THIS TIME. RESULTS AND CONCLUSIONS WILL BE PROVIDED IN THE FINAL REPORT.

Additional Manufacturer Narrative

THE DEVICE WAS NOT RETURNED FOR ANALYSIS. PRODUCTION RECORD AND CORRECTIVE AND PREVENTATIVE ACTIONS (CAPA) REVIEWS WERE PERFORMED AND REVEALED NO INDICATION OF A PRODUCT QUALITY ISSUE. ADDITIONALLY, A QUERY OF THE COMPLAINT HANDLING DATABASE FOR THE REPORTED LOT REVEALED THERE IS NO INDICATION OF A LOT SPECIFIC ISSUE. BASED ON THE INFORMATION PROVIDED, A DEFINITIVE CAUSE FOR THE REPORTED FAILURE TO ADVANCE COULD NOT BE DETERMINED. FACTORS THAT COULD CONTRIBUTE TO FAILURE TO ADVANCE INCLUDE, BUT ARE NOT LIMITED TO, PATIENT ANATOMICAL MORPHOLOGY, PATIENT DISEASE STATE, PRE-DILATATION STRATEGY, DEVICE PLACEMENT TECHNIQUE, INTERACTION WITH PREVIOUSLY PLACED STENTS OR ACCESSORY DEVICES. IN THIS CASE, IT IS POSSIBLE THE DEVICE MAY HAVE INTERACTED WITH THE MODERATELY CALCIFIED, MODERATELY TORTUOUS LESION DURING ADVANCEMENT, CAUSING THE REPORTED FAILURE TO ADVANCE; HOWEVER, BASED ON THE INFORMATION PROVIDED BY THE ACCOUNT AND WITHOUT THE DEVICE TO EXAMINE, THIS CANNOT BE CONFIRMED. BASED ON THE RESULTS OF THE COMPLAINT INVESTIGATION, THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO THE DESIGN, MANUFACTURE, OR LABELING OF THE DEVICE. D9 CORRECTION: DEVICE AVAILABLE FOR EVALUATION UPDATED FROM YES TO NO

Additional Manufacturer Narrative

A VISUAL AND DIMENSIONAL INSPECTION WAS PERFORMED ON THE RETURNED DEVICE. THE REPORTED FAILURE TO ADVANCE COULD NOT BE EVALUATED AS THE EXACT ANATOMICAL CONDITIONS ENCOUNTERED BY THE DEVICE USED DURING THE PROCEDURE COULD NOT BE REPLICATED IN THE TEST LABORATORY. PRODUCTION RECORD AND CORRECTIVE AND PREVENTATIVE ACTIONS (CAPA) REVIEWS WERE PERFORMED AND REVEALED NO INDICATION OF A PRODUCT QUALITY ISSUE. ADDITIONALLY, A QUERY OF THE COMPLAINT HANDLING DATABASE FOR THE REPORTED LOT REVEALED THERE IS NO INDICATION OF A LOT OF SPECIFIC ISSUE. BASED ON THE INFORMATION PROVIDED, A DEFINITIVE CAUSE FOR THE REPORTED FAILURE TO ADVANCE COULD NOT BE DETERMINED. FACTORS THAT COULD CONTRIBUTE TO FAILURE TO ADVANCE INCLUDE, BUT ARE NOT LIMITED TO, PATIENT ANATOMICAL MORPHOLOGY, PATIENT DISEASE STATE, PRE-DILATATION STRATEGY, DEVICE PLACEMENT TECHNIQUE, INTERACTION WITH PREVIOUSLY PLACED STENTS OR ACCESSORY DEVICES. IN THIS CASE, IT IS POSSIBLE THE DEVICE MAY HAVE INTERACTED WITH THE MODERATELY CALCIFIED, MODERATELY TORTUOUS LESION DURING ADVANCEMENT, CAUSING THE REPORTED FAILURE TO ADVANCE; HOWEVER, BASED ON THE INFORMATION PROVIDED BY THE ACCOUNT, THIS CANNOT BE CONFIRMED. BASED ON THE RESULTS OF THE COMPLAINT INVESTIGATION, THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO THE DESIGN, MANUFACTURE, OR LABELING OF THE DEVICE. D9 CORRECTION: DEVICE AVAILABLE FOR EVALUATION UPDATED TO YES.