inforMED
InjuryLSL

APTIMA COMBO 2 ASSAY

Received May 30, 2025 · Event occurred May 17, 2025

Report 2024800-2025-00029 · MDR key 22116412

Device

Generic name

Dna-reagents, Neisseria

Manufacturer

Hologic, Inc.

Catalog number

302661

Lot number

915709

Product problems

  • Off-Label Use
  • Off-Label Use

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

ON (B)(6) 2025, A CUSTOMER REPORTED TO HOLOGIC RECEIVING DISCREPANT RESULTS USING APTIMA COMBO 2 ASSAY (AC2) MASTER LOT 915709 ON PANTHER PLUS INSTRUMENT. SAMPLE ID# (B)(6) INITIALLY RESULTED CT POSITIVE & GC POSITIVE ON AC2 (B)(4) ON PANTHER PLUS INSTRUMENT SN (B)(6). THE SAME SAMPLE WAS RETESTED AND RESULTED CT NEGATIVE & GC NEGATIVE ON AC2 (B)(4) ON PANTHER PLUS INSTRUMENT SN (B)(6). CUSTOMER REPORTED THE RETESTED CT NEGATIVE & GC NEGATIVE RESULT TO THE PATIENT. PER APTIMA COMBO 2 ASSAY PACKAGE INSERT ¿THE FIRST VALID RESULT FOR EACH ANALYTE IS THE RESULT THAT SHOULD BE REPORTED¿. CUSTOMER WAS INFORMED THAT NOT REPORTING OUT THE FIRST VALID RESULT IS CONSIDERED OFF-LABEL. CUSTOMER NOTED THAT THEY ARE AWARE BUT PER THEIR LAB SOP, ALL DUAL POSITIVE SAMPLES WITH RELATIVE LIGHT VALUES (RLUS) <1900 ARE TO BE REPEATED AND IF THE RETEST IS NEGATIVE, THEY REPORT OUT THE NEGATIVE RESULT. NO PATIENT TREATMENT INFORMATION WAS PROVIDED BY THE CUSTOMER. HOLOGIC HAS NOT LEARNED OF ANY ADVERSE PATIENT OUTCOMES DUE TO THIS EVENT.

Additional Manufacturer Narrative

HOLOGIC TECHNICAL SUPPORT (TS) REVIEWED ALL OF WORKLISTS AND NOTED NO HARDWARE OR REAGENT PREPARATION ISSUES. SUSPECTED POTENTIAL SAMPLE MISHANDLING OR LOW TARGET SAMPLES. HOLOGIC PERFORMED A RISK ASSESSMENT AND NOTED NO PRODUCT IMPACT. HOLOGIC HAS NOT BEEN INFORMED OF ANY ADVERSE PATIENT OUTCOMES RELATED TO THIS ISSUE.