inforMED
MalfunctionKIJ

CS META AUG REAMER HOUSING HW

Received May 28, 2025 · Event occurred May 6, 2025

Report 1818910-2025-08603 · MDR key 22099882

Device

Generic name

Shoulder Instrument - Alignment Device

Catalog number

231770124

Lot number

N0042326Q

Product problems

  • Device-Device Incompatibility
  • Device-Device Incompatibility

Patient

NA · Male

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE DEVICE MALFUNCTION, DEVICE INTERACTION (2+ DEVICES): JAMMED/SEIZED, WILL NO LONGER BE REPORTED AS THERE HAS NOT BEEN A SERIOUS INJURY ASSOCIATED WITH THIS FAILURE WITHIN THE PAST 2 YEARS. PLEASE CONSIDER THIS THE LAST MANUFACTURER REPORT NUMBER SUBMITTED AGAINST THIS MALFUNCTION.

Additional Manufacturer Narrative

THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803 (AND/OR PART 4, AS APPLICABLE). THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem

IT WAS REPORTED THAT DISPOSABLE REAMER HEAD COLD WELLED ONTO THE REAMER SHAFT.

Additional Manufacturer Narrative

PRODUCT COMPLAINT #(B)(4). THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES JOINT RECONSTRUCTION, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES JOINT RECONSTRUCTION, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. H11 ADDITIONAL NARRATIVE: ADDED: D9, G1 (MANUFACTURING SITE NAME). IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative

PRODUCT COMPLAINT # (B)(4). INVESTIGATION SUMMARY : ACCORDING TO THE INFORMATION RECEIVED, " IT WAS REPORTED THAT DISPOSABLE REAMER HEAD COLD WELLED ONTO THE REAMER SHAFT." THE PRODUCT WAS RETURNED TO J&J MEDTECH ORTHOPAEDICS FOR EVALUATION. VISUAL INSPECTION OF THE RETURNED CS META AUG REAMER HOUSING HW REVEALS THAT IT ARRIVED ASSEMBLED WITH THE CS META AUG REAMER HEAD. A FUNCTIONAL TEST WAS PERFORMED MANUALLY AND WAS ABLE TO REPLICATE THE REPORTED JAMMED CONDITION DUE TO THE SUBCOMPONENTS WERE UNABLE TO DISENGAGE AFTER MULTIPLE ATTEMPTS WITH EXCESSIVE FORCES APPLIED. THE OBSERVED CONDITION OF THE DEVICE COULD BE DUE TO AN INTERNAL DAMAGE OF THE LOCKING MECHANISM. HOWEVER, DUE TO THE LOW FREQUENCY OF THIS OCCURRENCES AND THE INACCESSIBILITY OF THE INTERNAL COMPONENTS WITHOUT DESTRUCTION OF THE DEVICE A POTENTIAL CAUSE CANNOT BE ESTABLISHED. A DIMENSIONAL INSPECTION WAS NOT PERFORMED AS IT IS NOT APPLICABLE TO THE COMPLAINT CONDITION THE OVERALL COMPLAINT WAS CONFIRMED AS THE OBSERVED CONDITION OF THE CCS META AUG REAMER HOUSING HW WOULD HAVE CONTRIBUTED TO THE COMPLAINED DEVICE ISSUE. BASED ON THE INVESTIGATION FINDINGS, THE POTENTIAL CAUSE CANNOT BE ESTABLISHED, AND IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE REPORTED COMPLAINT CONDITION. AS PART OF J&J MEDTECH QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. DEVICE HISTORY LOT : THE PRODUCT INVESTIGATION FOUND NO EVIDENCE SUSPECTING AN ERROR IN THE MANUFACTURING OR MATERIAL THAT WOULD BE A CONTRIBUTING FACTOR IN THE REPORTED ALLEGATION(S). A MANUFACTURING RECORDS EVALUATION (MRE) WAS NOT PERFORMED.