Narrative

Description of Event or Problem

THE CUSTOMER COMPLAINED FALSE NEGATIVE REACTIONS OF A D POSITIVE CONTROL SAMPLE WITH CELL 1 OF BIOTESTCELL 3. CUSTOMER HAD SENT NEITHER THE COMPLAINED LOT OF BIOTESTCELL 3 NOR THE CONTROL.

Additional Manufacturer Narrative

SUMMARY: THE REVIEW OF THE BATCH RECORD DOCUMENTATION SHOWED NO IRREGULARITIES WHICH MIGHT HAVE AFFECTED NEGATIVELY THE QUALITY OF THE COMPLAINED LOT. DUE TO THE CONFIRMED COMPLAINT A RISK ANALYSIS WAS PERFORMED. ON THE BASIS OF THE RESULT OF THE RISK ANALYSIS WE DECIDED THAT NO MARKET RELATED ACTIVITIES ARE NECESSARY. A CAPA HAS BEEN INITIATED TO INTENSIFY THE EFFORTS TO FIND THE ROOT CAUSE FOR THE COMPLAINT.