inforMED
MalfunctionMVR

GEM MICROVASCULAR ANASTOMOTIC COUPLER

Received May 22, 2025 · Event occurred Apr 18, 2025

Report 1416980-2025-02993 · MDR key 22074941

Device

Generic name

Device,anastomotic,microvascular

Model number

GEM2754

Catalog number

511100300060

Lot number

SP25A06-2200836

Product problems

  • Material Twisted/Bent
  • Incomplete or Inadequate Connection
  • Material Twisted/Bent
  • Incomplete or Inadequate Connection

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

ADDITIONAL INFORMATION WAS ADDED TO D4, D9, H3, H4, H6, AND H11: H11: THE DEVICE WAS RECEIVED FOR EVALUATION. A VISUAL INSPECTION WAS PERFORMED, AND A MISALIGNMENT WAS VISIBLE. THE MISALIGNMENT WOULD NOT ALLOW THE COUPLER RINGS TO BE PROCESSED. THE REPORTED CONDITION WAS VERIFIED. THE CAUSE OF THE CONDITION COULD NOT BE DETERMINED. A BATCH REVIEW WAS CONDUCTED AND THERE WERE NO DEVIATIONS FOUND RELATED TO THIS REPORTED CONDITION DURING THE MANUFACTURE OF THIS LOT. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem

IT WAS REPORTED THAT THE PINS OF A 3.0MM COUPLER WERE CROOKED AND COULD NOT BE FULLY ATTACHED. THIS WAS OBSERVED DURING PRE-OPERATION SET UP. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Additional Manufacturer Narrative

SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.