inforMED
MalfunctionHTZ

2.0/2.4/2.7 PLATE CUTTER

Received May 20, 2025 · Event occurred Apr 29, 2025

Report 8030965-2025-05049 · MDR key 22056300

Device

Generic name

Instrument, Cutting, Orthopedic

Manufacturer

Synthes Gmbh

Catalog number

03.430.502

Lot number

KM967039

Product problems

  • Break
  • Break

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

PRODUCT COMPLAINT # (B)(4). DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem

IT WAS REPORTED THAT ON (B)(6) 2025 NEW INSTRUMENT CAME APART. THE EVENT OCCUR DURING STERILE PROCESSING. THE EVENT OCCUR DURING FIELD INSPECTION. THERE WAS NO PATIENT CONSEQUENCES.

Additional Manufacturer Narrative

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. H11 ADDITIONAL NARRATIVE: H3, H6: PHYSICAL DEVICE INVESTIGATION. THE PRODUCT WAS RETURNED TO J&J MEDTECH ORTHOPAEDICS FOR EVALUATION. VISUAL INSPECTION OF THE RETURNED DEVICE FOUND THE RASP FEATURE WAS NOT RETURNED WITHIN THE DEVICE, INDICATING THE COMPONENT FELL APART. EVEN THOUGH BREAKAGE WAS NOT IDENTIFIED, THE REPORTED CONDITION CAN BE CONFIRMED. NO OTHER ISSUE WAS OBSERVED. THIS ISSUE WAS BEEN ESCALATED AND CONFIRMED TO BE MANUFACTURING RELATED. IT WAS CONCLUDED THAT INCOMPLETE INSPECTION CRITERIA CONTRIBUTED TO THE ISSUE AS IT ALLOWED TO MUCH VARIABILITY IN RASP PLACEMENT. WHEN THEN THE RASP IS POORLY SEATED, IT LEAVES THE RASP SUSPECTABLE TO IMPACT AND FALLING OUT. A DIMENSIONAL INSPECTION WAS NOT PERFORMED SINCE IT WAS NOT APPLICABLE TO THE COMPLAINT CONDITION. A FUNCTIONAL TEST WAS NOT PERFORMED SINCE IT WAS NOT APPLICABLE TO THE COMPLAINT CONDITION. THE OVERALL COMPLAINT WAS CONFIRMED AS THE OBSERVED CONDITION OF THE 2.0/2.4/2.7 PLATE CUTTER WOULD HAVE CONTRIBUTED TO THE COMPLAINED DEVICE ISSUE. BASED ON THE INFORMATION CURRENTLY AVAILABLE, A PRODUCT ISSUE WAS IDENTIFIED DURING THE INVESTIGATION OF THE SAMPLE RECEIVED. THIS PRODUCT ISSUE WILL BE ADDRESSED THROUGH J&J MEDTECH ORTHOPAEDICS QUALITY SYSTEM. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. H3, H6: PHOTO INVESTIGATION THE PRODUCT WAS NOT RETURNED TO MEDTECH ORTHOPAEDICS; HOWEVER, A PHOTO WAS PROVIDED FOR REVIEW. SEE ATTACHMENT IMG_2931.JPG. THE PHOTO INVESTIGATION REVEALED THAT THE RASP FEATURE WAS OBSERVED UNATTACHED. EVEN THOUGH BREAKAGE WAS NOT IDENTIFIED, THE REPORTED CONDITION CAN BE CONFIRMED. THIS ISSUE WAS BEEN ESCALATED AND CONFIRMED TO BE MANUFACTURING RELATED. IT WAS CONCLUDED THAT INCOMPLETE INSPECTION CRITERIA CONTRIBUTED TO THE ISSUE AS IT ALLOWED TO MUCH VARIABILITY IN RASP PLACEMENT. WHEN THEN THE RASP IS POORLY SEATED, IT LEAVES THE RASP SUSPECTABLE TO IMPACT AND FALLING OUT. AS THE DEVICE WAS NOT RETURNED, AN AS-RECEIVED CONDITION COULD NOT BE ASSESSED, AND A DIMENSIONAL INSPECTION AND DOCUMENT/SPECIFICATION REVIEW WERE NOT COMPLETED. THE OVERALL COMPLAINT WAS CONFIRMED AS THE OBSERVED CONDITION OF THE 2.0/2.4/2.7 PLATE CUTTER WOULD CONTRIBUTE TO THE COMPLAINED DEVICE ISSUE. BASED ON THE INFORMATION CURRENTLY AVAILABLE, A PRODUCT ISSUE WAS IDENTIFIED DURING THE INVESTIGATION. THIS PRODUCT ISSUE WILL BE ADDRESSED THROUGH J&J MEDTECH ORTHOPAEDICS QUALITY SYSTEM. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. DEVICE HISTORY LOT: PART # 03.430.502, SYNTHES LOT # KM967039, SUPPLIER LOT # KM967039, RELEASE TO WAREHOUSE DATE: 10 DEC 2024, SUPPLIER: (B)(4). NO NCR'S GENERATED DURING PRODUCTION. DEVICE HISTORY REVIEW: REVIEW OF THE DEVICE HISTORY RECORD(S) SHOWED THAT THERE WERE NO ISSUES DURING THE MANUFACTURE OF THIS PRODUCT, AND ANY SUB-COMPONENTS, WHICH WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.