AMSCO 600
Received May 16, 2025 · Event occurred Feb 19, 2025
Report 3005899764-2025-00041 · MDR key 22036265
Device
Product problems
- Connection Problem
- Insufficient Device Problem Information
- Connection Problem
- Insufficient Device Problem Information
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
INVESTIGATION OF THE REPORTED EVENT IS IN PROGRESS; A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE. IT SHOULD BE NOTED THAT THIS SPECIFIC MODEL IS NOT SOLD IN THE UNITED STATES, SO IT IS NOT IN GUDID.
Description of Event or Problem
THE USER FACILITY REPORTED THAT AN ALARM OCCURRED, AND THE CYCLE ABORTED WHILE PROCESSING INSTRUMENTS IN THEIR AMSCO 600 STERILIZER. THE INSTRUMENTS WERE REPACKED AND REPROCESSED IN ANOTHER UNIT RESULTING IN A PROCEDURE DELAY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO REPORT OF INJURY.
Additional Manufacturer Narrative
THE USER FACILITY REPORTED THREE STERILIZERS SUBJECT OF THE REPORTED EVENT: (B)(6). FOLLOWING THE REPORTED EVENT, A STERIS ACCOUNT MANAGER LEARNED THE LOADS WERE INSIDE THE UNITS LONGER THAN EXPECTED DUE TO CONNECTIVITY ERROR ALARMS THAT PAUSED THE CYCLES, ATTRIBUTING TO PROCEDURE DELAYS. THE USER FACILITY'S THIRD-PARTY SERVICE PROVIDER ARRIVED ONSITE TO INSPECT THE AMSCO 600 STERILIZER SUBJECT OF THE REPORTED EVENT AND CONFIRMED THE HMI CABLE AND CONTROLLER CABLE WERE DISCONNECTED. THE SERVICE PROVIDER RECONNECTED THE CABLES, TESTED THE UNIT, CONFIRMED IT TO BE OPERATING ACCORDING TO SPECIFICATIONS, AND RETURNED IT TO SERVICE. THE AMSCO 600 STERILIZER OPERATOR MANUAL STATES, "TROUBLESHOOTING - CONNECTION ISSUE; HAVE BIOMED VERIFY THE CONNECTIONS TO THE DISPLAY OR CONTACT STERIS." A 3-YEAR COMPLAINT REVIEW INDICATES THIS TO BE AN ISOLATED EVENT. NO ADDITIONAL ISSUES HAVE BEEN REPORTED.