AMS 700 LGX? TENACIO? PUMP

Received May 16, 2025 · Event occurred Apr 29, 2025

Report 2124215-2025-31573 · MDR key 22035869

Device

Generic name

Prosthesis, Penis, Inflatable

Model number

72404456

Catalog number

72404456

Lot number

1100547772

Product problems

  • Deflation Problem
  • Unintended Deflation
  • Deflation Problem
  • Unintended Deflation

Patient

54 YR · Male

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

IT WAS REPORTED THAT THE PATIENT WITH AN INFLATABLE PENILE PROSTHESIS (IPP) EXPERIENCED DEFLATION DURING INTERCOURSE WITH TENACIO PUMP. A SURGICAL PROCEDURE WAS PERFORMED TO REMOVE AND REPLACE THE IPP. NO PATIENT COMPLICATIONS WERE REPORTED.

Additional Manufacturer Narrative

CONCOMITANT PRODUCT UDI FOR RESERVOIR IS (B)(4). UPON RECEIPT AT OUR QUALITY ASSURANCE LABORATORY, THIS INFLATABLE PENILE PROSTHESIS (IPP) UNDERWENT A THOROUGH ANALYSIS. THE PUMP WAS MICROSCOPICALLY INSPECTED, LEAK AND FUNCTIONALLY TESTED. THE PUMP PASSED VISUAL INSPECTION. NO DAMAGE OR ABNORMALITIES WERE FOUND. THE PUMP PASSED BOTH LEAKAGE AND FUNCTIONAL TEST. BASED ON THE INFORMATION AVAILABLE AND ANALYSIS RESULTS, THE REPORTED DEVICE ALLEGATION IS UNABLE TO BE CONFIRMED. THEREFORE, A CONCLUSION CODE OF CAUSE NOT ESTABLISHED WAS ASSIGNED TO THIS INVESTIGATION.

Description of Event or Problem

IT WAS REPORTED THAT THE PATIENT WITH AN INFLATABLE PENILE PROSTHESIS (IPP) EXPERIENCED DEFLATION DURING INTERCOURSE WITH TENACIO PUMP. A SURGICAL PROCEDURE WAS PERFORMED TO REMOVE AND REPLACE THE IPP. NO PATIENT COMPLICATIONS WERE REPORTED.