Received May 16, 2025 · Event occurred Feb 9, 2024

Report 3014312726-2024-00330 · MDR key 22035688

Device

Generic name

Sol-m 1ml Luer Lock Syringe W/o Needle (pp)

Model number

P180001PP

Lot number

04104028

Product problems

  • Fluid/Blood Leak
  • Fluid/Blood Leak

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

CUSTOMER REPORTED: NEEDLE LEAKED ALL THE BOTOX OUT.

Additional Manufacturer Narrative

THERE IS NO EVIDENCE OF QUALITY ISSUES ASSOCIATED WITH THIS COMPLAINT BASED ON THE RETURN SAMPLE AND ARCHIVE SAMPLE ANALYSIS. BASED ON THE DETERMINED RISK PRIORITY NUMBER, THE RESIDUAL RISK RELATED TO THE COMPLAINT IS CONSIDERED ACCEPTABLE. SOL MILLENNIUM CONTINUES TO MONITOR THIS KIND OF ISSUE. ADDITIONAL ASSESSMENT WILL BE TAKEN IF A SIGNIFICANT PATTERN EMERGES. UNCONFIRMED. THIS REPORT WAS INITIALLY SUBMITTED ON 10/15/2024. DUE TO ISSUES WITH DISPLAYING THE INITIAL SUBMISSION, THIS SUBMISSION CONTAINS BOTH INITIAL AND SUPPLEMENTAL INFORMATION.

Additional Manufacturer Narrative

THIS REPORT WHEN INITIALLY SUBMITTED ENCOUNTERED AN INTERNAL ISSUE POTENTIALLY RELATED TO DATABASE CONNECTIVITY. THIS ISSUE WAS IDENTIFIED WHILE IMPLEMENTING CAPA-57 WHICH WAS OPENED FOLLOWING AN FDA INSPECTION.

Remedial action

  • Notification