inforMED
MalfunctionQNQ

SOL-MILLENNIUM

Received May 16, 2025 · Event occurred Jan 26, 2024

Report 3014312726-2024-00192 · MDR key 22035476

Device

Generic name

Sol-m 1ml Luer Lock Syringe W/o Needle

Model number

P180001PP

Lot number

04105025

Product problems

  • Crack
  • Component Missing
  • Crack
  • Component Missing

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

"CUSTOMER REPORTED: WE RECEIVED A COMPLAINT FROM RIVM FOR ITEM P180001PP WITH BATCH 04105025. IT HAD SEVERAL DEFECTS: SYRINGE WITHOUT GRADUATION MARKS: 1 PIECE. SYRINGE WITHOUT PLUNGER: 1 PIECE. CLOSED PACKAGING WITHOUT SYRINGE: 1 PIECE. SYRINGE EXPANDED AND WITH FISSURES; FRACTURE CRACKS: 3 PIECES."

Additional Manufacturer Narrative

THERE IS NO EVIDENCE OF QUALITY ISSUES ASSOCIATED WITH THIS COMPLAINT BASED ON THE ARCHIVE SAMPLE ANALYSIS. BASED ON THE DETERMINED RISK PRIORITY NUMBER, THE RESIDUAL RISK RELATED TO THE COMPLAINT IS CONSIDERED ACCEPTABLE. SOL MILLENNIUM CONTINUES TO MONITOR THIS KIND OF ISSUE. ADDITIONAL ASSESSMENT WILL BE TAKEN IF A SIGNIFICANT PATTERN EMERGES. UNCONFIRMED. THIS REPORT WAS INITIALLY SUBMITTED ON 04/08/2024. DUE TO ISSUES WITH DISPLAYING THE INITIAL SUBMISSION, THIS SUBMISSION CONTAINS BOTH INITIAL AND SUPPLEMENTAL INFORMATION.

Additional Manufacturer Narrative

THIS REPORT WHEN INITIALLY SUBMITTED ENCOUNTERED AN INTERNAL ISSUE POTENTIALLY RELATED TO DATABASE CONNECTIVITY. THIS ISSUE WAS IDENTIFIED WHILE IMPLEMENTING CAPA-57 WHICH WAS OPENED FOLLOWING AN FDA INSPECTION.