TRANSLUMINAL BIPOLAR PACING PROBE
Received May 15, 2025 · Event occurred Apr 11, 2025
Report 2015691-2025-03855 · MDR key 22028110
Device
Generic name
Electrode, Pacemaker, Temporary
Manufacturer
Edwards Lifesciences PrModel number
PE074F5Lot number
65623390
Product problems
- Pacing Problem
- Pacing Problem
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
ONE SWAN GANZ PACING CATHETER WAS RETURNED FOR EVALUATION. CUSTOMER REPORT OF PACING ISSUE WAS CONFIRMED. CONTINUITY TESTING CONFIRMED A FULL OPEN CONDITION OF THE DISTAL CIRCUIT. THE PROXIMAL CIRCUIT WAS FOUND TO BE CONTINUOUS. A CUT DOWN OF THE CATHETER BODY WAS PERFORMED JUST PROXIMAL OF THE PROXIMAL ELECTRODE TO EXPOSE THE PACING LEADWIRES. IT WAS OBSERVED THAT THE DISTAL LEADWIRE WAS BROKEN NEAR THE DISTAL ELECTRODE. THE BALLOON INFLATED CLEAR AND CONCENTRIC WITH 1.3 CC AIR AND REMAINED INFLATED FOR MORE THAN 5 MINUTES WITHOUT LEAKAGE. NO VISIBLE DAMAGE OR DEFECT WAS OBSERVED FROM THE BALLOON, WINDINGS, CATHETER BODY AND RETURNED SYRINGE. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED AND DOCUMENTED THAT DEVICE MET ALL SPECIFICATIONS UPON DISTRIBUTION. AN ENGINEERING EVALUATION WAS INITIATED TO ASSESS FOR ANY MANUFACTURING-RELATED PROCESSES WHICH COULD BE CORRELATED TO THE COMPLAINT. BASED ON THE AVAILABLE INFORMATION THERE IS NO EVIDENCE THAT SUPPORTS OR CONFIRMS THE FAILURE MODE IS ASSOCIATED TO A MANUFACTURING/ DESIGN DEFECT. THEREFORE A ROOT CAUSE COULD NOT BE ESTABLISHED. AS PER MANUFACTURING PROCESS, A VERIFICATION OF THE NICKEL MAGNET BIFILAR FOR DELAMINATION IS PERFORMED. ALSO, THIS PROCEDURE PROVIDES INSTRUCTION FOR THE PROPER WAY TO INSERT THE BIFILAR WIRE INTO THE CATHETER TUBE. THE PROCEDURE INCLUDES INSTRUCTIONS TO PERFORM THE SOLDERING OF THE DISTAL ELECTRODES. THEN, THE PROCEDURE ESTABLISHED INSTRUCTIONS FOR THE PREPARATION OF THE PROXIMAL FILAMENT AND A CONTINUITY TEST FOR THE DISTAL AND PROXIMAL ELECTRODES. A FINAL CONTINUITY TEST IS PERFORMED TO THE ELECTRODES. THE INSTRUCTIONS FOR USE (IFU) PROVIDES THE FOLLOWING WARNINGS AND PRECAUTIONS: THIS CATHETER REQUIRES SPECIAL TECHNIQUES FOR INSERTION AND REMOVAL. ELECTRODE DISLODGEMENT MAY RESULT FROM PULLING THE CATHETER OUT THROUGH THE PERCUTANEOUS SHEATH. AVOID FORCEFUL WIPING OR STRETCHING OF THE CATHETER DURING TESTING AND CLEANING AS NOT TO BREAK THE ELECTRODE WIRE CIRCUITRY. SINCE PROPER FUNCTIONING OF THE PACING CATHETER DEPENDS ON THE ELECTRICAL CONTINUITY OF ITS ELECTRODES AND INTERNAL WIRES, CARE SHOULD BE EXERCISED WHEN HANDLING THE CATHETER.
Additional Manufacturer Narrative
THE DEVICE EVALUATION IS ANTICIPATED. HOWEVER, THE COMPLAINT CANNOT BE CONFIRMED WITHOUT THE COMPLETION OF A PRODUCT EVALUATION. A SUPPLEMENTAL REPORT WILL BE FORTHCOMING WHEN THE INVESTIGATION IS COMPLETED. THE DEVICE HISTORY RECORD REVIEW HAS NOT YET BEEN COMPLETED. COMPLAINT HISTORIES FOR ALL REPORTED EVENTS ARE REVIEWED AGAINST TRENDING CONTROL LIMITS ON A MONTHLY BASIS AND ANY EXCURSIONS ABOVE THE CONTROL LIMITS ARE ASSESSED AND DOCUMENTED AS A PART OF THE MONTHLY REVIEW.
Description of Event or Problem
AS REPORTED DURING USE AND AFTER INSERTION IN THE PATIENT THIS SWAN GANZ PACING CATHETER WAS NOT ENERGIZED AND DID NOT PACE FROM THE BEGINNING WHEN IT WAS CONNECTED TO THE EXTERNAL PACEMAKER. WHEN THIS CATHETER WAS REPLACED, THE ISSUE WAS SOLVED. THERE WAS NO ALLEGATION OF PATIENT INJURY. THE DEVICE WILL BE AVAILABLE FOR PRODUCT EVALUATION.