inforMED
MalfunctionCEK

TOTAL PROTEIN GEN.2

Received May 14, 2025 · Event occurred Apr 21, 2025

Report 1823260-2025-01481 · MDR key 22017820

Device

Generic name

Biuret Total Protein

Manufacturer

Roche Diagnostics

Catalog number

08058652190

Lot number

838819

Product problems

  • Low Test Results
  • Non Reproducible Results
  • Low Test Results
  • Non Reproducible Results

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

THERE WAS AN ALLEGATION OF QUESTIONABLE TOTAL PROTEIN GEN.2 RESULTS FOR 1 PATIENT SAMPLE ON A COBAS C 703 ANALYTICAL UNIT. THE INITIAL RESULT WAS 0.119 G/DL. THE REPEATED RESULT WAS 7.81 G/DL. THE REPEATED RESULT MATCHED THE PATIENT'S CLINICAL HISTORY.

Additional Manufacturer Narrative

THE ANALYZER'S SERIAL NUMBER IS (B)(6). THE CALIBRATION AND QC WERE ACCEPTABLE. THE ALARM TRACE SHOWED SEVERAL "ABNORMAL ASPIRATION ALARMS". A HARDWARE PERFORMANCE CHECK SHOWED THAT ROTOR A WAS OUT OF SPECIFICATION. THE INVESTIGATION IS ONGOING.

Additional Manufacturer Narrative

THE INVESTIGATION DETERMINED THAT THE ROOT CAUSE WAS CONSISTENT WITH A HARDWARE ISSUE CAUSING THE TRIGLYCERIDES FOR ROTOR A TO BE OUT OF SPECIFICATION. THE INVESTIGATION DID NOT IDENTIFY A SPECIFIC CAUSE OF THE EVENT.