A&E MEDICAL / ROTATING SURGICAL PUNCHES

Received May 13, 2025 · Event occurred Apr 17, 2025

Report 0002242056-2025-00011 · MDR key 22011684

Device

Generic name

Punch, Surgical

Catalog number

080-401

Lot number

22660

Product problems

  • Contamination /Decontamination Problem
  • Contamination /Decontamination Problem

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. THE FOLLOWING SECTIONS WERE UPDATED: B4; B5; D2; D4; D9; G1; G3; G6; H1; H2; H3; H4; H6; H10. THE REPORTED EVENT HAS BEEN CONFIRMED. A VISUAL INSPECTION WAS CONDUCTED ON THE RETURNED PRODUCT. THE RETURNED PACKED DEVICE WAS STILL SEALED. VISUAL INSPECTION OF THE PACKAGING SHOWED ONE PARTICLE NEAR THE TIP OF THE DEVICE. A REVIEW OF THE DEVICE HISTORY RECORDS IDENTIFIED NO DEVIATIONS OR ANOMALIES DURING MANUFACTURING. THE REPORTED EVENT DID NOT OCCUR IN AN OPERATING ROOM OR AS PART OF A MEDICAL PROCEDURE; MEDICAL RECORDS ARE NOT AVAILABLE FOR REVIEW. THE ROOT CAUSE OF THE REPORTED ISSUE IS ATTRIBUTED TO A MANUFACTURING PACKAGING ISSUE ADDRESSED IN A PREVIOUS CORRECTIVE ACTION. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Description of Event or Problem

NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

Additional Manufacturer Narrative

(B)(4). MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT. ASSOCIATED REPORTS ARE AVAILABLE FOR THE APPLICABLE CONCOMITANT PRODUCTS BELOW IN THE D10 NARRATIVE. D10: CONCOMITANT MEDICAL PRODUCTS, PART (LOT): ASSOCIATED PRODUCT INFORMATION: - 080-351 (20669). - 080-351 (20669). - 080-351 (10333). - 080-351 (13141). - 080-351 (22660). G2: FOREIGN - THE EVENT OCCURRED IN TAIWAN (R.O.C.). THE CUSTOMER HAS INDICATED THAT THE PRODUCT IS IN THE PROCESS OF BEING RETURNED TO ZIMMER BIOMET FOR INVESTIGATION. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED. ATTEMPTS HAVE BEEN MADE TO GATHER ALL PRODUCT IDENTIFICATION INFORMATION, AND NO FURTHER INFORMATION HAS BEEN PROVIDED. IF ANY FURTHER INFORMATION IS FOUND THAT WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Description of Event or Problem

IT WAS REPORTED THAT A HOSPITAL INVENTORY CLERK FOUND UNIDENTIFIED OBJECTS IN THE PACKAGING DURING THEIR INSPECTION PROCESS. THERE WAS NO PATIENT INVOLVEMENT OR IMPACT, AS THE OBJECTS WERE IDENTIFIED OUTSIDE OF A CLINICAL SETTING. THE CORRECT SURGICAL TECHNIQUE WAS USED WHEN INSPECTING THE PRODUCT PACKAGING. IT WAS REPORTED THAT NO FURTHER INFORMATION IS AVAILABLE.