SECHRIST INDUSTRIES BLENDER 3000-71318

Received May 12, 2025 · Event occurred Apr 27, 2025

Report 22001948 · MDR key 22001948

Device

Generic name

Mixer, Breathing Gases, Anesthesia Inhalation

Product problems

  • Activation, Positioning or Separation Problem
  • Material Integrity Problem

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

PATIENT ECMO [EXTRACORPOREAL MEMBRANE OXYGENATION] CANNULA WERE THE SAME DARK COLOR ON A SWEEP OF 1 AND 80%. GOOD CATCH BY [REDACTED] THE BEDSIDE NURSE. ECMO COORDINATOR CHANGED THE O2 [OXYGEN] LINE TO THE PORTABLE O2 TANK AND THE COLOR OF THE CANNULA CHANGED TO A BRIGHT RED. ATTEMPTED TO BURP THE BLENDER BUT THE BLENDER STILL DID NOT WORK. PATIENT SPO2 [PERIPHERAL OXYGEN SATURATION] WAS STABLE DURING THIS PERIOD.