MalfunctionBZR
SECHRIST INDUSTRIES BLENDER 3000-71318
Received May 12, 2025 · Event occurred Apr 27, 2025
Report 22001948 · MDR key 22001948
Device
Generic name
Mixer, Breathing Gases, Anesthesia Inhalation
Manufacturer
Sechrist Industries, IncProduct problems
- Activation, Positioning or Separation Problem
- Material Integrity Problem
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
PATIENT ECMO [EXTRACORPOREAL MEMBRANE OXYGENATION] CANNULA WERE THE SAME DARK COLOR ON A SWEEP OF 1 AND 80%. GOOD CATCH BY [REDACTED] THE BEDSIDE NURSE. ECMO COORDINATOR CHANGED THE O2 [OXYGEN] LINE TO THE PORTABLE O2 TANK AND THE COLOR OF THE CANNULA CHANGED TO A BRIGHT RED. ATTEMPTED TO BURP THE BLENDER BUT THE BLENDER STILL DID NOT WORK. PATIENT SPO2 [PERIPHERAL OXYGEN SATURATION] WAS STABLE DURING THIS PERIOD.