KETONE
Received May 7, 2025 · Event occurred Apr 10, 2025
Report 1000113657-2025-00160 · MDR key 21977783
Device
Generic name
Nitroprusside, Ketones (urinary, Non-quant.)
Manufacturer
Trividia Health IncModel number
STRIP, WALGREENSKETONE 50CT #496636Lot number
AC726
Product problems
- No Device Output
- No Device Output
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
INTERNAL REPORT REFERENCE NUMBER: (B)(4). KETONE TEST STRIPS WERE NOT RETURNED FOR EVALUATION. RETENTION TESTING WAS PERFORMED USING TEST STRIPS FROM THE SAME LOT. RETENTION STRIP LOT TESTED WITHIN SPECIFICATIONS. MOST LIKELY UNDERLYING ROOT CAUSE: (B)(4): USER HAD AN INACCURATE REFERENCE. NOTE: MANUFACTURER MADE SEVERAL ATTEMPTS TO CONTACT CUSTOMER TO ENSURE THE INITIAL CONCERN HAS BEEN RESOLVED- UNABLE TO ESTABLISH CONTACT WITH CUSTOMER.
Description of Event or Problem
CONSUMER REPORTED COMPLAINT FOR THE TRUEPLUS KETONE TEST STRIPS, STATING THAT THEY WERE NOT CHANGING COLOR. THE PACKAGE WAS NOT OPEN OR DAMAGED WHEN RECEIVED BY THE CUSTOMER. THE CUSTOMER IS USING THE PRODUCT FOR THE FIRST TIME OUT OF THIS PACKAGE. THE CUSTOMER IS USING PROPER TESTING TECHNIQUES. THE CUSTOMER WAS NOT WILLING TO PERFORM A URINE TEST DURING THE CALL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. NO MEDICAL ATTENTION RELATED TO THE USE OF THE PRODUCT WAS REPORTED.
Remedial action
- Other