STATSTRIP GLUCOSE HOSPITAL METER SYSTEM
Received May 2, 2025 · Event occurred Mar 31, 2025
Report 1219029-2025-00002 · MDR key 21953028
Device
Generic name
Blood Glucose Meter For Near-patient Testing
Manufacturer
Nova Biomedical CorporationModel number
1.75Catalog number
53398
Lot number
0323206249
Product problems
- High Test Results
- High Test Results
Patient
36 YR · Female
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
THE COMPLAINANT REPORTED "CONFLICTING RESULTS BETWEEN 2 GLUCOMETERS AND PATIENT'S DEXCOM. NURSE ALSO SAID THE PATIENT'S DEXCOM WAS "OUT OF WACK" AS WHEN JUICE WAS GIVEN THE DEXCOM READING DID NOT CHANGE."
Additional Manufacturer Narrative
THE METER IN QUESTION WAS NOT RETURNED FOR ANALYSIS. THE METER WAS PASSING QC BEFORE AND AFTER THE SUSPECT RESULT. INCREASED FINGERSTICKS WERE DONE DUE TO POSSIBLE ERRONEOUS TEST RESULT. WITHOUT THE RETURNED PRODUCT A DETAILED INVESTIGATION ON THE FAILED PRODUCT COULD NOT BE PERFORMED, AND THE REPORTED CONDITION COULD NOT BE CONFIRMED. THIS COMPLAINT FILE SHALL BE CLOSED AS UNCONFIRMED AT THIS TIME. A DHR REVIEW WAS PERFORMED. NO ABNORMALITIES OR CONCERNS WERE OBSERVED. THE DHR INDICATED THE RELEASED PRODUCT MET ALL SPECIFICATIONS. A COMPARISON TO A REFERENCE ANALYZER WAS NOT PERFORMED. NO FURTHER ACTION IS RECOMMENDED AT THIS TIME. NOVA BIOMEDICAL WILL CONTINUE TO MONITOR FOR THIS OR SIMILAR EVENTS.