EARLYVUE VS30 VITALS MONITOR
Received Apr 30, 2025 · Event occurred Apr 1, 2025
Report 9610816-2025-000373 · MDR key 21935545
Device
Generic name
Earlyvue Vs30 Vitals Monitor
Manufacturer
Philips Medizin Systeme Böblingen GmbhModel number
863380Catalog number
863380
Product problems
- Incorrect Measurement
- Incorrect Measurement
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
A BIOMEDICAL EQUIPMENT TECHNICIAN (BET) VISITED ONSITE AND CONFIRMED THE REPORTED PROBLEM. THE BET DETERMINED THE ISSUE WAS CAUSED DUE TO A DEFECTIVE SPO2 CABLE. THE CUSTOMER WILL ORDER THE REPLACEMENT SPO2 CABLE THEMSELVES. THE INVESTIGATION CONCLUDES THAT NO FURTHER ACTION IS REQUIRED AT THIS TIME. IF ADDITIONAL INFORMATION IS RECEIVED THE COMPLAINT FILE WILL BE REOPENED.
Description of Event or Problem
IT WAS REPORTED THAT, WHILE THE DEVICE IS ON, THE EARLY VUE VS30 VITALS MONITOR READS A FALSE SPO2 SATURATION RATE. THE DEVICE WAS IN CLINICAL USE AT THE TIME OF THE EVENT. NO ADVERSE EVENT OR PATIENT HARM WAS REPORTED.
Additional Manufacturer Narrative
PHILIPS IS IN THE PROCESS OF OBTAINING ADDITIONAL INFORMATION CONCERNING THIS EVENT AND THE COMPLAINT IS STILL UNDER INVESTIGATION. A FINAL REPORT WILL BE SUBMITTED ONCE THE INVESTIGATION IS COMPLETE.