C-FLEX POLAR HEAD POSITNER- DEVICE ONLY
Received Apr 29, 2025 · Event occurred Apr 22, 2025
Report 3010216206-2025-00008 · MDR key 21926622
Device
Generic name
C-flex Head Positioning System
Manufacturer
Baxter Healthcare CorporationModel number
A-70701Catalog number
A-70701
Lot number
N/A
Product problems
- Positioning Failure
- Positioning Failure
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THE HEAD POSITIONER IS DESIGNED TO POSITION AND SUPPORT THE PATIENT¿S HEAD IN A VARIETY OF SURGICAL PROCEDURES INCLUDING, BUT NOT LIMITED TO SPINE SURGERY. THESE DEVICES ARE INTENDED TO BE USED BY HEALTHCARE PROFESSIONALS WITHIN THE OPERATING ROOM SETTING. THE CFLEX HEAD POSITIONER WAS RECEIVED BY BAXTER. UPON INSPECTION, IT WAS DETERMINED THAT THE TOP PIVOT WAS NOT HOLDING AND NEEDED ADJUSTMENT. THE TOP PIVOT WAS ADJUSTED AND INNE CAM SYSTEM WAS REPLACED. BASED ON THIS INFORMATION, NO FURTHER ACTIONS ARE REQUIRED AT THIS TIME. ALTHOUGH THERE WAS NO REPORTED INJURY WITH THIS EVENT, IF THE REPORT OF A CFLEX HEAD POSITIONER HAVING AN UNINTENDED MOVEMENT WERE TO RECUR, IT COULD POTENTIALLY CAUSE SERIOUS INJURY OR DEATH. THEREFORE BAXTER IS REPORTING THIS EVENT.
Description of Event or Problem
BAXTER RECEIVED A REPORT STATING THAT CFLEX POLAR HEAD POSITIONER JOINT WAS NOT HOLDING CORRECT AMOUNT OF WEIGHT. THE DEVICE WAS LOCATED AT THE ACCOUNT AND OCCURRED BEFORE USE. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.